6541-4-538
3/8˝ IM Drill
6541-4-800
T-Handle Driver
6541-4-516
5/16˝ IM Rod
6541-2-600
Tibial Alignment Jig IM
6541-4-801
Universal Driver
Figure 36
Instrument Bar
25
Varus/Valgus Alignment
>
Assemble the appropriate Tibial Resection Guide
(left, right or Universal Resection Guide) on the
Tibial Adjustment Housing .
Note: The Tibial Adjustment Housing is available in 3°
slope and 0° slope (optional) .
>
Attach the assembly onto the mounting bar by
pressing the bronze wheel on the Tibial Adjustment
Housing . Attach the Universal Alignment Handle
to the Tibial Resection Guide and slide a Universal
Alignment Rod through the handle for sagittal
assessment .
>
When alignment is confirmed, the Universal
Alignment Handle should be centered over the ankle .
Tibial
Preparation
0º slope 6541-2-704
3º slope 6541-2-705
Tibial Adjustment Housing
Right 6541-2-700
Left 6541-2-701
Tibial Resection Guide
6541-4-602
Universal Alignment Rods
6541-1-721
Universal Resection Guide
( Tibial Resection
Guide shown)
6541-4-806
Universal Alignment Handle