![Stryker Triathlon Tritanium Surgical Manualline Download Page 41](http://html.mh-extra.com/html/stryker/triathlon-tritanium/triathlon-tritanium_surgical-manualline_1386067041.webp)
Impactor
Pad
Locking
Lever
6541-4-810
Impaction Handle
6541-4-807
Femoral Impactor/Extractor
Figure 57
Figure 56
Instrument Bar
39
Component Implantation
Tibial Implantation
Locking onto the Impactor Pad
Note: It is advisable to confirm at this stage if the surgeon
plans to implant the Baseplate with or without cement .
If Cementless:
>
Connect the Tibial Baseplate Impactor/Extractor to the
Impaction Handle
>
Connect the Tibial Baseplate Impactor/Extractor to the
Triathlon Tritanium Baseplate over the Impactor Pad and
lock the lever
>
The Triathlon Tritanium Tibial Baseplate is packaged
together with an Impactor Pad as shown .
Impactor
Pad
5536-B-X00 –
Baseplate – See Catalog
6541-4-901 – Impactor Pad
Triathlon Tritanium Tibial Baseplate with
Impactor Pad
6541-4-805
Baseplate Impactor/Extractor