![Stryker Triathlon Tritanium Surgical Manualline Download Page 25](http://html.mh-extra.com/html/stryker/triathlon-tritanium/triathlon-tritanium_surgical-manualline_1386067025.webp)
Fixation
Pin
(6541-2-611 Tibial
Alignment Proximal Rod EM)
Locking
Switch
Right 6541-2-700
Left 6541-2-701
Tibial Resection Guide
Instrument Bar
23
Tibial
Preparation
0º slope 6541-2-704
3º slope 6541-2-705
Tibial Adjustment Housing
6541-2-609
Tibial Alignment Ankle Clamp EM
6541-2-611
Tibial Alignment Proximal Rod EM
6541-2-610
Tibial Alignment Distal Assembly EM
6541-2-611E
Express Proximal Rod EM
Rotational Alignment
>
Rotate the entire assembly to ensure that the base of
the assembly is aligned with the center of the ankle .
>
The center of the ankle is generally in line with the
second metatarsal .
>
Once alignment is confirmed, set the bronze locking
switch on the Distal Assembly to the locked position .
Figure 32