0210-100-700 Rev-T
EN
www.stryker.com
7
Environmental
Conditions:
Operation
Storage and Transportation
Temperature Limitation:
Humidity Limitation:
Atmospheric Pressure
Limitation:
Guidance and manufacturer’s declaration - electromagnetic emissions
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is intended for use in the electromagnetic environment specified below.
The customer or the user of the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series uses RF energy only
for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is suitable for use in
all establishments, including domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not Applicable
Guidance and manufacturer’s declaration - electromagnetic immunity
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is intended for use in the electromagnetic environment specified below.
The customer or the user of the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series,
including cables, than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
d=[1.2]
√
P
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
d=[1.2]
√
P
80 MHz to 800 MHz
d=[2.3]
√
P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m)
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey
a
, should be less than the compliance level in each frequency
range
b
.
Interference may occur in the vicinity of equipment marked with the following
symbol:
(Non-ionizing electromagnetic radiation)
NOTE 1: At 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is used exceeds the applicable RF compliance level above, the
InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
0000004059, Rev. T Effective Date: Sep 26, 2016 2:47:57 PM