Maintenance Procedures
Chapter 4
51
11. Using a 1/2" wrench, remove the bolts/shoulder spacers(4)/locknuts (
R
) holding the torsion
levers (
S
) to the lower part of the two counter-levers (
M
).
12. Remove the two torsion levers (
S
) from both ends of the torque shaft (
T
) by first removing the
two locknuts (
U
) using a 1/2" wrench, and then the two screws (
V
) using a 3/16" Allen key.
N
N
O
O
T
T
E
E
Apply grease on the portion of the torsion levers that inserts into the nylon bushings (
W
) before
replacing them.
When reassembling, screw down tightly the two screws (
V
) holding the torsion levers to the torque
shaft before tightening the two locknuts (
U
).
13. Using a 1/2" wrench, remove the bolt/nut (
F
) holding the caster to the swing arms and remove
the caster.
14. Reverse the above steps to install the replacement swing arm assembly. Read carefully the
following note to properly reinstall the pedals.
N
N
O
O
T
T
E
E
Be sure to position the pedal horizontally (neutral) when inserting its shaft into the sockets of the
locking levers (
E
, figure 4.17, page 47) and the 5th wheel activation lever (
J
).
When inserting the pedal shaft into the sockets, make sure that the locking lever is vertically
positioned and that the upper flat part of the 5th wheel activation lever is parallel to the floor (bring
the 5th wheel near the floor by exerting a pressure on it to ease this operation).
15.
Verify the 5th wheel mechanism for proper operation before returning the bed to service.