Maintenance Procedures
Chapter 4
33
HEAD SECTION
•
BED WITHOUT OPTIONAL CPR
Required Tools:
Phillips Screwdriver
5/32" Allen key
Needle Nose pliers
Ratchet Tie-Down Strap
1/2" Wrench
OG2 Grease
Procedure:
1. Raise the bed to high position, lock the casters and bring the mattress support to horizontal
position.
2. Remove the head board and raise the Fowler to high position.
3. Unplug the power cord from the wall receptacle.
4. Remove both head safety siderails. Refer to the "Head Siderail Assembly" replacement
procedure, page 16.
The four bolts used to mount a siderail assembly cannot be reused because their Scotch-Grip
coating is less efficient once they have been tightened and removed thereafter. They
must
be
replaced with new identical bolts.
5. Using a Phillips screwdriver, remove the two mattress side retainers (
A
) and save them for
the replacement head section.
6. Plug the bed power cord and bring the Fowler back to horizontal position.
7. Using needle nose pliers, remove the Rue ring/nylon washers(3)/clevis pin (
E
, fig. 4.12,
page 37) hooking up the head actuator tube to the head section lever arms.
N
N
O
O
T
T
E
E
Apply grease on the clevis pin and the nylon washers before hooking up the actuator tube to the
head section.
Figure 4.11D
TÊTE
A
SEAT SECTION
HEAD
END
HEAD SECTION
D
D
C
B
W
W
A
A
R
R
N
N
I
I
N
N
G
G