The
GO
BED Maintenance Manual
44
4.15 CPR MICRO SWITCH
MICRO SWITCH LOCATED UNDER THE HEAD SECTION
Required Tools:
Ratchet Tie-Down Strap
7/16" Wrench
1/2" Wrench
Phillips Screwdriver
Procedure:
1. Raise the bed to the high position and lock the casters.
2. Raise the head section to the high position and attach it securely to the foot section using
bungee cords to prevent it from dropping off when the upper part of the small locking lever is
disassembled from the head section at step 7.
3. Unplug the power cord from the wall receptacle.
4. Remove the head panel.
5. Using a 7/16" wrench, remove the nut (
E
) holding the activation rod (
F
) to the activation
lever, disengage rod and lay it down.
6. Disconnect the two wires (
J
, fig. 4.11F, page 36) from the micro switch. Note the wire
connecting positions.
7. Using a 1/2" wrench, remove the bolt/shoulder spacers(4)/washers(4)/locknut (
K
, fig. 4.11F,
page 36) linking the upper part of the small locking lever to head section. Disengage the
lever and lay the assembly down. Ensure that the head section is securely attached to the
bed before disassembling the small locking lever.
8. Using a Phillips screwdriver, remove the two screws (
A
) holding the micro switch (
B
) to the
head section long lever. Keep the mounting plate (
C
). Remove the defective micro switch.
9. Reverse the above steps to install the replacement micro switch.
10. Verify the CPR mechanism for proper operation before returning the bed to service.
Figure 4.15
C
B
A
D