Maintenance Procedures
Chapter 4
19
4.2 MEMBRANE REPLACEMENT - FOOT BOARD CONTROL PANEL
N
N
O
O
T
T
E
E
The membrane comes with the control panel plate, already affixed to it.
Required Tools:
Phillips Screwdriver
Procedure:
1. Raise the bed to the high position and lock the casters.
2. Unplug the power cord from the wall receptacle.
3. Remove the foot board and lay it on a workbench
4. Using a Phillips screwdriver, remove the 7 screws (
A
) holding the support cover (
B
) to the
inner face of the foot board and remove the cover.
5. Using a Phillips screwdriver, remove the screw (
C
) holding the membrane support to the foot
board.
6. Disconnect the membrane cable (
E
) from the board connector (
D
).
7. Remove the control panel plate (
F
).
N
N
O
O
T
T
E
E
On some beds, the membrane cable may be stuck to the inner face of the board using a small
transfer tape.
8. Reverse the above steps to install the replacement membrane.
N
N
O
O
T
T
E
E
Be sure to connect the new membrane cable correctly. The side of the membrane connector
bearing the inscription "1" must face the black wire of the foot board connector.
9.
Verify all the foot board controls for proper operation before returning the bed to service.
A
B
C
D
E
F
Figure 4.2
Summary of Contents for GOBED FL17E