AUTO CONTOUR MICRO SWITCH REPLACEMENT
Required Tools:
Phillips Screwdriver
#1 Phillips screwdriver
Procedure:
1. Fully raise the bed and apply the brakes.
2. Raise the thigh section and fold the foot section back toward the head end of the bed.
1
st
micro switch:
This switch tells the control board that the Knee Gatch has reached the preset
angle for the
Auto Contour
position.
3. Remove the cable wires from the switch. Note their location so they will be connected
correctly to the new switch.
4. Since the support has oblong holes to adjust its
position, we will position the thigh section in such a
way that the correct mounting adjustment will be easy
to obtain. Using the Knee Gatch down control, slowly
lower the thigh section until only a light contact
remains between the activator (D), attached to the
knee section, and the switch push button (E). The
replacement procedure will start from this point (see
opposite illustration).
5. Unplug the bed power cord from the wall outlet.
6. Using a Phillips screwdriver, remove the two screws
(F) holding the support to the frame.
N
N
O
O
T
T
E
E
Mount the support at the same position than the one mentioned at step 4.
7. Press the two switch clips to remove from the support.
8. Reverse the above steps to install the new micro switch.
Figure 4.3G
A
B
C
D
E
F
G
H
Summary of Contents for FL26E