![Stryker FL26E Maintenance Manual Download Page 31](http://html.mh-extra.com/html/stryker/fl26e/fl26e_maintenance-manual_1385904031.webp)
Maintenance
Procedures
Chapter 4
31
N
N
O
O
T
T
E
E
Make sure that the support and the molded nut holes are aligned before screwing in the bolts. If
resistance is felt, it means that the holes are not aligned.
9. Remove the defective actuator.
10. Reverse the above steps to install the new actuator. Carefully read the following caution
before hooking up the actuator to the Hi-Lo lever.
The course of the new actuator must be adjusted prior to hooking it to the Hi-Lo lever. An
improper adjustment can damage the Hi-Lo mechanism.
10. To adjust the new actuator, proceed as follows:
A. Once the new actuator cable is connected, plug the bed power cord and press the bed
down control until the actuator stops. This will be the lower limit of the actuator course.
B. Link the actuator to the Hi-Lo lever.
C. The alignment jigs still in position, raise and lower completely the bed to ensure that
the lower limit has been preserved.
C
C
A
A
U
U
T
T
I
I
O
O
N
N