Maintenance Manual
16
No Knee Gatch up or down motion when:
•
the siderail command is used
•
the foot end command is used
•
Is the Knee Gatch lockout activated (LED
(padlock) on) ? If so, deactivate it.
•
Is the total lockout activated (LED (padlock)
on) ? If so, deactivate it.
•
Is the cable of the siderail control panel
properly connected to the bed connector
under the mattress support?
•
Check points of the “No power to bed”
problem described above.
No
Auto Contour
motion
•
Is the Knee Gatch or total lockout activated
(LED (padlock) on) ? If so, deactivate it.
•
Check and replace if needed (see page 37)
the two Auto Contour limit switches.
Improper operation of the CPR positioning:
Knee Gatch does not lower and/or the Fowler
actuator does not reset.
•
Check and replace if needed (see pages
38) the two CPR limit switches.
Nurse call signal does not reach the nurse
desk.
•
Check points of the “No power to bed”
problem described above.
•
Is the nurse call cable properly connected
to the bed 37 pin connector located at the
head end of the bed as well as in the nurse
call wall outlet?
•
Check the wall outlet.