![Stryker FL26E Maintenance Manual Download Page 18](http://html.mh-extra.com/html/stryker/fl26e/fl26e_maintenance-manual_1385904018.webp)
Maintenance Manual
18
FOOT END SIDERAIL ASSEMBLY REPLACEMENT
Required Tools:
1/2" Socket Wrench (w/6" Extension)
Torque Wrench
Procedure:
1. Fully raise the bed and apply the brakes. Raise the siderail needing repair.
2. Fully raise the Knee Gatch and unplug the bed power cord from the wall receptacle.
3. Lift and fold the foot section back toward the head end of the bed.
4. Using a 1/2" socket wrench with a 6" extension, remove the four thread-rolling bolts (A)
holding the siderail assembly to the support and remove the assembly. Support the siderail
assembly while removing the last bolts.
N
N
O
O
T
T
E
E
Once removed, the thread-rolling bolts must be first screwed in manually before tightening them
to a 130 lbf in (14.7 N m) torque using a torque wrench.
5. Reverse the above steps to install the new siderail assembly.
6. Test the siderail for proper operation before returning the bed to service.
A
B
D
E
F
G
H
I
Figure 4.1B
C
Summary of Contents for FL26E
Page 2: ......
Page 45: ...Appendix A Circuit Diagrams 45 ...