![Stryker FL26E Maintenance Manual Download Page 15](http://html.mh-extra.com/html/stryker/fl26e/fl26e_maintenance-manual_1385904015.webp)
Troubleshooting
Chapter 3
15
3. TROUBLESHOOTING
Please consult the following troubleshooting guide before calling the Technical Service department
(see section 1.2).
3.1 TROUBLESHOOTING GUIDE
PROBLEM/FAILURE
WHAT TO CHECK
No power to bed
•
Is the bed power switch at the head end of
the bed turned on?
•
Is the power cord connected to the power
connector and plugged into the wall outlet?
•
Is the power cord severed? Replace if
needed.
•
Are the two fuses inside the power
connector still operational (see page 34)
•
Verify power at wall outlet.
No bed up or down motion when:
•
the siderail command is used
•
the foot end command is used
•
Is the total lockout activated (LED (padlock)
on) ? If so, deactivate it.
•
Is the cable of the siderail control panel
properly connected to the bed receptacle
under the mattress support?
•
Check points of the “No power to bed”
problem described above.
No Fowler up or down motion when:
•
the siderail command is used
•
the foot end command is used
•
Is the Fowler lockout activated (LED
(padlock) on) ? If so, deactivate it.
•
Is the total lockout activated (LED (padlock)
on) ? If so, deactivate it.
•
Is the cable of the siderail control panel
properly connected to the bed connector
under the mattress support?
•
Check points of the “No power to bed”
problem described above.
The Fowler does not
fully
raise
This situation happens when the CPR handle
is used to partly lower the Fowler. The use of
the CPR mechanism for this purpose creates a
situation where the Fowler motor is out of sync
with the actual position of the Fowler. To
correct the situation:
•
Completely lower the Fowler using the
CPR handle or the Fowler down control to
enable the Fowler motor to reset itself.
Refer to the "CPR Emergency Release"
section of the Operations Manual for more
information.
Summary of Contents for FL26E
Page 2: ......
Page 45: ...Appendix A Circuit Diagrams 45 ...