10
Basic Maintenance
Cleaning the X8000
Caution
Unplug the X8000 before cleaning the
unit.
1. As needed, clean the external surfaces of the X8000
using a cloth or sponge dampened with a mild
detergent or disinfectant.
2. Clean and maintain the light cable according to the
manufacturer’s instructions.
Caution
Do not use any abrasive cleaners. Do not
allow any liquid to drip into the unit.
Caution
Do not sterilize or immerse the X8000.
Caring for the Bulb Module
e X8000 uses a Xenon bulb, which has a guaranteed
life of 500 hours when used properly. Always follow these
guidelines to ensure maximum bulb life:
1. If the bulb is touched, clean the bulb face with
alcohol and a cotton swab. Dirt or oil on the bulb
face will cause the bulb to heat unevenly and fail.
2. Do not power on and o
ff
the bulb in rapid
succession. Allow the bulb to run for at le ast
fi
ve
minutes once it has been powered on. Failure to do
so can cause the bulb to rapidly darken and fail.
Caution
e X8000 has been designed to start the
bulb under most conditions, even when
it is hot. However, if the bulb does not
start within 10 seconds a
er the X8000
has powered on, turn the unit o
ff
and
wait at least
fi
ve minutes for the bulb to
cool before restarting. Further attempts
to start the bulb can damage the bulb and
possibly the internal circuitry.
Replacing the Bulb Module
Replace the bulb module when the LCD indicates 500 hours
or when the bulb no longer su
ffi
ciently illuminates the
surgical site. If possible, replace the bulb module between
surgical procedures.
Warning
During operation, the bulb and the
housing around the bulb may be hot.
Wait at least three minutes for the bulb
to cool before handling it.
To replace the bulb module,
1. Power down the X8000 and wait at least three
minutes for the bulb to cool.
2. Open the bulb door.
3. Rotate the bulb module handle to the right, grasp it,
and withdraw the bulb.
Warning
Do not reach inside the bulb door for
any reason other than replacing the bulb
module. Touching parts other than the
bulb module may cause burns or product
damage.
Warning
Do not touch any part of the bulb module
except for the handle.
e module may be
very hot and cause burns.
4. Insert the new bulb module (Stryker part number
220-201-000) along the guide rails until it is fully
seated on the mating connectors.
5. Rotate the handle to the le
to secure the bulb
module within the light source.
6. Close and latch the bulb door.
7. Power on the unit.
8. Power on the bulb (if not already on) and verify that
the LCD displays the bulb hours as “0.”
Note
Using a non-Stryker replacement bulb
module will result in no light output.
Caution
Do not operate the X8000 with a burned-
out bulb or with no bulb installed.