![Stryker 2131 Operation Manual Download Page 29](http://html.mh-extra.com/html/stryker/2131/2131_operation-manual_1385805029.webp)
En
gl
is
h
2131-009-005 REV A
1-25
SYSTEM MESSAGE INDICATOR LIGHTS
Note
• The total lockout mechanism blocks the control panel from the siderails, footboard, pendant, touch screen and
head panel (located between the ZOOM
®
Option handles) .
• The brake system can be activated during a Total Lockout but it cannot be removed at any moment during a Total
Lockout .
• The Total Lockout does not affect the Nurse Call (optional), Scale System (optional) and Bed Exit Detection
(optional) functions .
• The parameters of the lock mechanism will be saved if there is a power outage and will start from their original
state when the power returns .
Touch Screen
The Nurse Interface can be activated by pressing one of the function buttons or by pressing on the screen itself . After
2 minutes of inactivity, the LCD intensity will lower to about 10% of its maximal level .
If a submenu is opened and no further action is taken by the user, the LCD will shut off completely after 4 minutes (4
minutes = 2 minutes at 100% int 2 minutes at 10% intensity) .
Footboard Operation Guide
This indicator shows that the bed is no longer plugged into a wall socket and
that its power comes from the battery .
This indicator shows that the bed is actually plugged into a power outlet and
that the battery is recharging itself . It is highly recommended not to unplug
the bed as long as this LED is on .
This indicator shows that the brakes are not set, which means that the bed
might move if it is pushed or pulled . This LED can be seen as a reminder
to the personnel that the brake needs to be set before leaving a patient
unattended .
This indicator turns on when the power of the battery is at a low level . When
it does turn on, the bed should be plugged in a wall socket to allow the
battery to be recharged .
This indicator shows when the bed needs repairs or maintenance in regards
to its proper functioning . Most of the time the Call Maintenance indicator
will be illuminated when there is a problem with the electronic part of the
bed system . In this case the appropriate maintenance personnel should be
called .
This indicator shows that a part of the bed (Patient Control Only) is locked
or that a total lockout (All Bed Controls) is activated . Refer to
for
the Lockouts menu .
Bed Unplugged:
Bed Charging:
Brake Not Set:
Battery Low:
Call Maintenance:
Locks Enabled:
Summary of Contents for 2131
Page 2: ......
Page 4: ......
Page 40: ...English 1 36 2131 009 005 REV A www stryker com Return To Table of Contents Notes ...
Page 69: ...English www stryker com 2131 009 005 REV A 1 65 Return To Table of Contents Notes ...
Page 102: ...English 1 98 2131 009 005 REV A www stryker com Return To Table of Contents Notes ...
Page 138: ...Deutsch 2 36 2131 009 005 REV A www stryker com Zurück zum Inhaltsverzeichnis Notizen ...
Page 167: ...Deutsch www stryker com 2131 009 005 REV A 2 65 Zurück zum Inhaltsverzeichnis Notizen ...
Page 200: ...Deutsch 2 98 2131 009 005 REV A www stryker com Zurück zum Inhaltsverzeichnis Notizen ...
Page 236: ...Italiano 3 36 2131 009 005 REV A www stryker com Ritorna all indice Note ...
Page 265: ...Italiano www stryker com 2131 009 005 REV A 3 65 Ritorna all indice Note ...
Page 298: ...Italiano 3 98 2131 009 005 REV A www stryker com Ritorna all indice Note ...
Page 432: ...Eλληικά 5 36 2131 009 005 REV A www stryker com Επιστροφή στον πίνακα περιεχομένων Σημειώσεις ...
Page 461: ...Eλληικά www stryker com 2131 009 005 REV A 5 65 Επιστροφή στον πίνακα περιεχομένων Σημειώσεις ...
Page 494: ...Eλληικά Σημειώσεις ...
Page 495: ......