Unitom-2
Instruction Manual
21
Replace the cooling water in the Recirculation Cooling Unit at
least once a month.
Press emergency stop.
Open the protection guard.
Remove the red protection caps from each of the two grease
nipples, located as indicated with a white circle in the illustration.
Lubricate in the two grease nipples with the grease gun and
move the spindle/cutting motor up and down by hand, until
grease comes out at the shaft. Use water-repelling grease, e.g.
Shell Albida Grease LX.
Clean superfluous grease away and mount the protection caps
at the grease nipples.
Release the emergency stop.
Monthly service
Replacing the Cooling Water
Lubrication of Bearings for Tilting
Frame