6 Servicing
ASTODIA Instructions for Use
21
6 Servicing
6.1 Preventative servicing
ASTODIA does not require any preventative servicing.
No service or maintenance shall be carried out while in use with a
patient.
6.2 Periodic Safety Inspection
At least every 24 months - and/or in accordance with the applicable regional and
national regulations – it is necessary to check the ASTODIA Diaphanoscope for
safety reasons and to document it. The check consists of the following tests:
•
Visual inspection
•
Check the function of the yellow and the red LED
•
Check the functioning of the reduction of the maximum light intensity and
the automatic switching off.
•
Regionally and nationally required examinations of medical equipment
(e.g. electrical safety of the ASTODIA charger)
All tests of the periodic safety inspection have to be passed!
If one of the tests does not pass, the ASTODIA has to be sent to the local dealer
or to STIHLER ELECTRONIC GmbH for repair
(see Chapter 1.9 Return of a used
.
6.2.1 Visual inspection
Check the following components for completeness and visible damages before
each application.
•
Charger
•
Housing of the control unit
•
Control panel
•
Hand piece (Especially the plug of the cable)
6.2.2 Check the function of the yellow and the red LED
•
Switch on the ASTODIA on the control unit. The red and yellow LEDs each
has to flash three times. The red LED must light up at its lowest level.
•
Actuate the shift-key for the LED-colours. The yellow LED must light up at its
lowest level.
•
Check, if it is possible to change the light intensity by actuating the keys, see
Chapter 4.2 Starting.
•
Check, if the covering glasses of the LEDs are still clear and light-transmitting.