ASTODIA Instructions for Use
6 Servicing
22
6.2.3 Check the functioning of the reduction of the maximum light intensity
and the automatic switching off
•
Switch on the ASTODIA on the control unit and set the maximum light intensity
-
see Chapter4.2 Starting
•
Watch the LED and measure
I. the time, until the ASTODIA automatically reduces the maximum light in-
tensity by one level.
This time must be 2 m/- 10 s.
II. the time, until the ASTODIA automatically switches off completely.
This time must be 5 m/- 30 s.
6.2.4 Electrical Test
The charger has to be checked once a year according to the applicable national
safety regulations.
Summary of Contents for ASTODIA DIA100
Page 27: ......