9 Guidelines and manufacturer’s declaration
ASTODIA Instructions for Use
25
9 Guidelines and manufacturer’s declaration
Guidance and manufacturer's declaration - electromagnetic emissions
ASTODIA is intended for use in the electromagnetic environment specified below. The customer or user of the ASTODIA
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11/EN 55011
Group 1
ASTODIA uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause interference in nearby electronic equipment.
RF emissions
CISPR 11/EN 55011
Class A
The EMISSIONS characteristics of this equipment make it
suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might
not offer adequate protection to radio-frequency com-
munication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Harmonic emissions IEC/EN
61000-3-2
Class A
Voltage fluctuations / flicker
emissions
IEC/EN 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The ASTODIA is intended for use in the electromagnetic environment specified below.
The customer or user of the ASTODIA should assure that it is used in such an environment.
Immunity test
Test level
Compliance
level
Electromagnetic environment - guid-
ance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8
kV, ± 15 kV air
in compliance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/bursts
IEC/EN 61000-4-4
± 2 kV 100 kHz
repetition frequen-
cy
in compliance
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Surge
IEC/EN 61000-4-5
± 0.5 kV, ± 1 kV
Line-to-line
± 0.5 kV, ± 1 kV, ±
2 kV
Line- to-ground
in compliance
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Voltage dips IEC/EN
61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles
Single phase: at 0°
in compliance
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the ASTODIA requires
continued operation during power mains
interruptions, it is recommended that the
ASTODIA be powered from an uninter-
ruptible power supply or a battery.
Voltage interruptions
IEC/EN 61000-4-11
0 % U
T
; 250/300
cycle
in compliance
Rated power fre-
quency
magnetic fields
IEC/EN 61000-4-8
30 A/m
50 Hz or 60 Hz
in compliance
Power frequency magnetic fields should
be at levels characteristic of a typical
commercial or hospital environment.
NOTE U
T
is the AC mains voltage prior to application of the test level.