5-4
122997-341
Operator Manual
Cycle Programming
5.3.1 Pre-Wash
1.
Press
CHANGE VALUES
touch pad. Pre-Wash screen appears:
CYCLE 10
HTW
PRE-WASH
T=MM:SS
__indicates flashing position.
2.
To enter Pre-Wash phase time, press
CURSOR (left or right)
touch pad to select position and
VALUE (up or down)
touch
pad to select desired number (0-9). Phase time is input as min-
utes and seconds within a range of 0-49 minutes and 0-99 sec-
onds.
5.3.2 Alkaline Wash
1.
Press
CHANGE VALUES
touch pad. Alkaline Wash phase
values screen appears:
CYCLE 10
SP=140.0F
ALK.WASH
T=MM:SS
__indicates flashing position.
NOTE: If treatment time is set to 0, treatment is bypassed and
question is not displayed.
2.
To enter Wash temperature setpoint and phase time, press
CURSOR (left or right)
to select position and
VALUE (up or
down)
touch pad to select desired number (0-9). Temperature
setpoint is input as any number value to 1/10 of a degree,
between 70 and 185°F (22 and 85°C). Phase time is input as
minutes and seconds within a range of 0-49 minutes and
0-99 seconds.
3.
Press
CHANGE VALUES.
Alkaline wash Temperature Guaran-
tee option screen appears:
GUAR. ALK.WAS TEMP?
NO
__indicates flashing position.
4.
Press
VALUE (up or down)
touch pad to toggle between YES or
NO answer. Selecting NO starts Wash phase time countdown at
beginning of phase. Selecting YES guarantees that Wash phase
time counts down only when solution temperature is equal to or
greater than programmed set point.
NOTE: With Temperature Guarantee feature enabled, phase
time does not count down unless setpoint is reached. It is
important that setpoint be an attainable value. Guaranteed
temperature is shown by a G before setpoint (GSP instead of
SP). Wash chamber returned to drain is shown by an R before
setpoint (RGSP or RSP).
CHANGE
VALUES
CURSOR
VALUE
CHANGE
VALUES
CURSOR
VALUE
CHANGE
VALUES
VALUE