3-12
122997-341
Operator Manual
Component Identification
3.15 Drain Discharge
Cooldown with Side
Tank and Temperature
Guarantee
A separate tank is typically mounted along service side of washer to
temporarily retain and cool all drain discharges (see Figure 3-12).
NOTE: Location of drain discharge side tank may change depending
on other options ordered with washer.
During draining function of each phase, solution/water in sump is
automatically pumped to side tank where cold tap water is added.
Cold water remains on until water temperature in side tank reaches
programmed set point. Once cooled to set temperature, side tank
gravity drains to building drain system.
Figure 3-12. Drain Discharge Side Tank (Typical)
Drain Discharge Side
Tank