19
A regulation module inside the electrolyser
means that there may be an offset in time
between the signal emitted by the chlorine
analyser (STÉREDOX) and the electrolyser's
start (or stop) of production. Offset of 0 to 2
minutes.
Measurement < Regulation threshold
Case of STÉREDOX: "Regulation" LED is flashing
% Electrolyser Production = change from 0% to the percentage set
Measurement ≥ Regulation threshold
Case of STÉREDOX: "Regulation" LED is off.
% Electrolyser Production = change from 0%,
DEVICE REMAINS SWITCHED ON
D0rP DIAGNOSIS appears.
4.8 Displaying the diagnoses
After a certain operating time, a diagnosis appears. This diagnosis indicates that the electrode
needs to be checked.
See table below:
CODE
Diagnosis
Possible causes
Solutions
D001
This analysis
appears when the
device recognises
poor conductivity
between the
electrodes. It then
displays ‘---‘ on the
chlorine production
display.
The cell is disconnected
or not connected
properly, or oxidised
Check the connection
The electrodes are
scaled up
Clean the electrodes
(ACIDULOR
recommended cleaning
fluid)
The amount of salt in
the water is too low
Check the salt level 5g/l
D0033
After 6,000 hours of
operation of the
device, this analysis
appears.
It tells you that the
electrode needs to be
checked or replaced.
To cancel this analysis,
press on the A and F keys
at the same time.
D004
Flow rate sensor
fault
No flow,
sensor jammed,
valves closed
Check the flow of
water at the sensor, if
necessary
D0rP
The Stéredox box is
not in active
regulation, the
electrolyser is not
requested and the
production is at 0%
This diagnosis has just been established to justify this change
of production to 0% to the user, there is no malfunction.