16
DISPLAy FAULT MoDES
LED display reads
“A3”
• lamp life expired - countdown is at “0” days
• press reset button for a deferred alarm, replace UV lamp
LED display is
blank
• controller is in lamp failure mode
• power system down, allowing it to reset itself; apply power in order to
confirm that the controller is able to power lamp
• check to see if there is sufficient power to the UV system
Low UV level
displayed on
screen
• test water supply to see if water quality meets recommended
parameter limits
• clean quartz sleeve and sensor eye
LED flashing “A2”
and then back to
UV level
• low UV alarm deferral has been activated
• UV level has dropped below 50% and the audible alarm has been
muted by pressing the reset switch and holding it for 5 seconds
• this audible alarm deferral will only last 12 hours
Dose Flow Chart: