background image

34 

STEINDORFF

®

 

 

PROBLEM 

REASON 

SOLUTION 

PAGE 

II. Mechanical Part: 

1. The coarse focus 
knob is hard to run 

The tension adjustment collar is too 
tight 

Loose properly 

2. The image can’t stay 
on the focal when 
observation 

The tension adjustment collar is too 
loose 

Tighten properly 

III. Electric Part: 

1. The lamp can’t light 

No power supply 

Check the power cord, and connect 
them exactly 

the installation of the bulb is wrong 

Install the bulb correctly 

The bulb burn out 

Change a new bulb 

The bulb burns out in a 
high frequency 

Not use the specified lamp 

Use the required lamp  

1.

 

The height of the 
brightness is not 
enough  

Not use a appointed lamp 

Use an appointed lamp 

The brightness adjustment knob is 
used wrong 

Adjust the brightness adjustment 
knob in a correct way 

2.

 

The light glimpse 

The bulb is going to spoil 

Change the bulb 

The power cord have a poor contact 

Check the power cord, and connect 
them exactly 

. Viewing tube 

The two eyes’ field of 
view is different 

The interpupillary distance is not 
correct 

Adjust the interpupillary distance 

10 

The diopter is not right 

Adjust the diopter 

10 

Not adapter to the microscope 
observation 

When observing, do not stare at the 
specimen but at the whole field of 
view, or move the eyes away to see 
other things, then back into the 
objective  

 

. Microscope video 

1. The image is 
unfocused 

Focus incorrectly 

Adjusting the focus system, make 
the double reticle and the specimen 
distinctly to see 

10 

There is faintness 
around the image 

It is a inherent character of the 
achromatic objective 

The problem is unavoidable if you 
used an achromatic objective 

 

3.The indoor window or 
the fluorescence lamp 
develop 

The extra light entered into the 
eyepiece and viewfinder is reflected 

Cover up the eyepiece and the 
viewfinder of the microscope 
illumination system 

 

 
 
 

Summary of Contents for NYMCS-1701

Page 1: ...ual This manual is written for Inverted Fluorescence Microscope NYMCS 1701 For safety and for keeping the best performance making you familiar with the instrument entirely it is strongly recommended t...

Page 2: ...llation 5 2 2 Installment steps 6 3 Adjustment Operation 11 3 1 Lamp adjustment for fluorescence observation 17 3 2 Reflected observing illumination adjustment 17 3 3 Imaging Operation 19 4 Outfit 32...

Page 3: ...e setting of the voltage switch and the frequency switch on the rear of the power supply unit 7 Always use the power cord provided and make sure that the main switch is moved to O OFF before connectin...

Page 4: ...mpt to use organic solvents to clean the non optical component of the equipment To clean these use a lint free soft cloth lightly moistened with a diluted neutral detergent 3 Do not disassemble any pa...

Page 5: ...ontrast slider Lamp Reflected fluorescence units Condenser Illumination bracket Eyepiece Stage inserted plate Stage Body Nosepiece Objectives 4 10 20 40 Fluorescence power supplier Protection barrier...

Page 6: ...steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with the supplied hexagon wrench When changing the components you will need it again Condens...

Page 7: ...the setscrew and insert the Viewing Head into the body correctly screw down with bolt 2 2 2 Installing eyepiece fig 2 Insert the eyepiece into tube until they are against 2 2 3 Installing condenser s...

Page 8: ...it needn t to adjust in the use process If the ring is not in the center you could adjust by the centering bolt The 10X 20X light ring is worked with the 10X 20X phase contrast objective while the ope...

Page 9: ...used in the inversed microscope is very sensitivity about dust Do cover all the unused holes with turret dust caps to prevent the dust and contamination entering inside When operating use the low magn...

Page 10: ...ly unit and make sure the cord is correctly connected Make sure that the main switch of the power supply is set to O OFF before connecting cables 3 Connect the power cord connector into connector on t...

Page 11: ...fortable adjustment 1 Installing the stage lengthen splint First Screw the fixed bolt on to the splint then mount it on to the stage from right or left below screwing down it until it stay hard 2 Inst...

Page 12: ...t be re ignited for about 10mimutes that is until the mercury vapor inside it has cooled down and condenser to liquid Ensure that the hour counter is reset to 000 00 after replacement of the burner An...

Page 13: ...dicator pointed by the spot on the eyepiece holder shows the scale of the interpupillar distance The range of the interpupillar distance 48 75mm 3 1 5 Mounting Auxiliary Stage fig 4 1 When using mecha...

Page 14: ...ve the objective from the revolving nosepiece position and engage this position in the light path 5 Pull out the field iris diaphragm lever to close the iris diaphragm and push in the aperture iris di...

Page 15: ...making sure that the lamp housing has cooled down A mercury burner seals high pressure gas inside If the burner is used beyond its service life stress may accumulate inside the burner and in the worst...

Page 16: ...diaphragm lever out until the diaphragm comes into the smallest state 3 Use the hexangular wrench to adjust the two field iris diaphragm centering screws alternately to move the image of the diaphrag...

Page 17: ...Selecting the appropriate color filters according your need it became more effective to observe or photography the specimen Especially we suggest using the LBD color filter which can compensate more n...

Page 18: ...usted properly before leaving factory How to adjust tension of the coarse focusing knob Turn the tension adjustment collar While revolving at the direction of the arrow in the figure the tension of th...

Page 19: ...ight ring center and the phase contrast center when changing the objective If having departure you ought to center again If the light ring is centering incorrectly you will fail to obtain the best vie...

Page 20: ...n the control board WELCOME should appear on the LCD screen After 3 seconds the microscope system will be in the preview in the real time automatically when you can snap Adjust microscope s definition...

Page 21: ...ig17 01 Shown on the upper right corner is the Nos of photos that can be taken or the video recording time 0000 means in the current setup mode it still can shoot 0000pcs photos 00 00 00 means in curr...

Page 22: ...press the left key to select Yes and confirm by the menu button If you press No by menu button the photo remains stored on the LCD Microscope head itself 3 3 4 Function menu select With the help of th...

Page 23: ...ess menu button again to save your selection Exit from the setting up after confirmation the model comes back to the snap model Fig 18 02 On video model press menu button as shown the Size then press...

Page 24: ...comes back to the snap model Fig 20 Setup Setting up each item On snap model press button as shown the Size then press the down key to enter the Setup Then press the right key to select as Default Da...

Page 25: ...ey to select Date press the right key once again to select On Photo Off Photo Set Date by up down keys Press menu button to confirm selecting On Photo the present date and time will be indicated on th...

Page 26: ...the SD card inserted Attention Even the protected content will be deleted when formatting cannot be resumed again Select Cancel and press menu button the model comes back to the snap model F AE On mea...

Page 27: ...g on LCD screen On playback mode press menu button to find Delete function press right key and select Current or All use up down keys to delete single photo video or all photos video Press menu button...

Page 28: ...d confirm by the menu button Fig 21 03 Meantime the screen become blue Now we can open the installed driver and find PCCam on the desktop of PC Choose option to preview the real time imaging or take p...

Page 29: ...28 STEINDORFF Preview Take video...

Page 30: ...ion 800 600 of microscope LCD screen so through AV outputting TV can show a sharper picture than on microscope screen Also the microscope shoots pictures as 2560 1920 pixels when playback on TV or PC...

Page 31: ...there are 5colors available green blue black white and red It used to select among colors of the self definition graphics facilitating the clear indication under different color materials If you pres...

Page 32: ...ion Optical system Infinite optical system NYMCS 1701 Reflected light source Excitation units excitation Dichroic mirror Barrier filter Blue Excitation BP460 490 DM500 BA520 Green Excitation BP510 550...

Page 33: ...mm Filter 45mm blue green and ground glass Focusing Coaxial coarse and fine adjustment vertical objectives movement Coarse stroke 37 7mm per rotation Fine stroke 0 2mm per rotation Operation condition...

Page 34: ...solution problems Image is not sharp The contrast is not high The detail is not clear Don t obtain the phase contrast effect The nosepiece is not in the center of the light path Ensure the nosepiece i...

Page 35: ...ulb is going to spoil Change the bulb 3 The power cord have a poor contact Check the power cord and connect them exactly 6 Viewing tube The two eyes field of view is different The interpupillary dista...

Reviews: