3. Adjustment and operation
3.1 Lamp adjustment for fluorescence observation
3.1.1 Connecting power
Set the main switch of the power supply unit to “I” (ON). It will stabilize in 5 to 10 minutes after ignition.
●
Some mercury burners may not ignite the first time the power is turned on due to variance in production,
and the safety mechanism in the starter in such a case. If this occurs, set the main switch to “1” (ON),
then press the starter reset switch on the front panel of the power supply and between 1 to 4 seconds are
required for igniting the burner. Repeat as necessary.
●
To avoid shortening the burner life, do not turn the burner off within 15 minutes after ignition.
● The burner cannot be re-ignited for about 10mimutes, that is, until the mercury vapor inside
it has cooled down and condenser to liquid.
● Ensure that the hour counter is reset to “000.00” after replacement of the burner. And you
can insert a thin object such as a mechanical pencil tip into the reset hole on the front panel
of the power supply unit to press the internal switch
3.1.2 Function of button (fig 1)
①
Hour counter
②
Ammeter
。
③
Excitation button
④
Start reset button
⑤
Voltage switch
3.1.3 Adjusting the diopter (fig 2)
1.
Look into the right ocular by your right eye, then
revolving the coarse focus knob to focus on the
specimen.
2.
Then use your left eye to look into the left ocular.
If the image is not sharp, just use the diopter
adjustment ring
①
to adjust please.
★
There are ±5 diopter in the adjustment ring.
The number which the reticle on the eyepiece
holder pointed is your eye’s diopter graduation.
Fig 1
1
2
3
①
②
④
⑤
③
on
①
off
①