background image

Multi Switch

Owner's Manual for the mo-Vis Multi Switch

Summary of Contents for IDM-MS

Page 1: ...Multi Switch Owner s Manual for the mo Vis Multi Switch...

Page 2: ...oll Free 800 806 1225 info stealthproducts com MDSS GmbH Schiffgraben 41 30175 Hannover Germany General Read and understand all instructions prior to the use of this product Failure to adhere to instr...

Page 3: ...and understand all safety aspects of the devices mentioned In order for our products to be used successfully you must read and understand all instructions and warnings maintain our products according...

Page 4: ...ree at 800 965 9229 Always keep the operating instructions in a safe place so they may be referenced as necessary All information pictures illustrations and specifications are based on the product inf...

Page 5: ...uct without written approval from Stealth Products LLC Any lack or alteration of serial number where applicable will automatically void this warranty Stealth Products LLC is liable for replacement par...

Page 6: ...Accessories 2 3 8 1 Multi Switch Package 2 8 2 Mounting the Multi Switch 3 9 0 Technical Data 4 13 9 1 USB Power Supply 4 9 2 Input 4 9 3 Output 4 9 4 mo Vis Configurator Software 5 9 5 Output Operati...

Page 7: ...mbly of product to end user Stealth Products LLC is not liable for user death or injury Testing 6 3 Initial setup and driving should be done in an open area free of obstacles until the user is fully c...

Page 8: ...3 5mm stereo input jack allowing a mechanical switch or proximity sensor input connection two 3 5mm stereo output jacks allowing up to four separate outputs with splitter cables sold separately a LED...

Page 9: ...eparately are compatible with the Multi Switch These are Optional Splitter Cable The 3 5mm stereo to 3 5mm mono splitter sold separately is required to convert a single output to a dual output Product...

Page 10: ...TICE Ensure the user has unobstructed visual contact with the LEDs on the front of the Multi Switch in order to check its status Although every activation is able to trigger an auditory signal it is n...

Page 11: ...el currently selected and shifts to select the desired channel A separate green LED on the face of the device indicates the device status When used with a splitter cable sold separately the Multi Swit...

Page 12: ...r or metal stick that approach the sensor The larger the surface of the sensor the more sensitive it will be the 24mm being the most sensitive The sensor is most sensitive at its front in the center I...

Page 13: ...click activate the input again to select and activate the output The output will react The output will remain closed as long as the input is closed or the output will remain locked until it is manuall...

Page 14: ...e If too much time elapsed without selecting and activating an output the selection process will stop and no output will be activated This is a built in feature designed to allow any unwanted input to...

Page 15: ...en working in this mode the output will be closed when the input is closed This means that when you break the lock your output will react as well Timed Open The output will be locked in the selected o...

Page 16: ...button on the front of the switch box and the system will store that position as the open reference When the calibration is successful a continuous beep will sound If the calibration has been unsucce...

Page 17: ...er level profile access This user can clear one or all of the errors from the fault log In the case of error codes please contact mo Vis to learn about the required actions Should a problem persist af...

Page 18: ...imity mode only 10 100 50 Proximity switch sensitivity higher value more sensitivity Reduce this value in case of false positives Output Sequence Tip Tip Select sequence Green Tip Yellow Tip Green Rin...

Page 19: ...longer Switched Output will toggle between open and closed Close Time 20 60000 200 Time in ms output will remain closed timed output only Lock No See 9 5 Output Operation Mode Selection Lock Mode p 8...

Page 20: ...5V Power Consumption 14 mA Input Mechanical switch Closed 200kOhm max Open 150kOhm min mo Vis proximity sensor 12mm or 24mm Output Max 60V 75mA Sensor Cable Length 120cm 3 93 ft Relevant Testing EN12...

Page 21: ...location for the Multi Switch Make sure user has visual contact with indication LEDs on the front of the Multi Switch Step Two Carefully remove the protective backing of one hook and loop patch apply...

Page 22: ...the user 3 Have the user test all possibilities of the button proximity sensor Are they well positioned for the user s needs Can they safely operate them with the least effort 4 Has the user tested al...

Page 23: ...th Cleaning 12 2 use a soft damp cloth to clean the device Ensure that all cleaners are approved for finished steel aluminum plastic and upholstered surfaces CAUTION Always check all mounting hardware...

Page 24: ...roducts LLC info stealthproducts com www stealthproducts com 800 965 9229 512 715 9995 104 John Kelly Drive Burnet TX 78611 P120D657 mo Vis Multi Switch Owners Manual pdf Jan 20 2021 2021 Stealth Prod...