66
As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice.
201.031
Date of Issue: 06 November 2018
11.2.1.2.1 Suggested General Settings
Figure 11.3 General Settings
Press the Settings button to open the Settings dialogue. The important settings for this walk-through are:
Test Duration; 10 seconds
Test Continuously: On, interval 0.
Start Testing Automatically: On
Stop Testing When Clean: Off
Simulate Test: On
NOTE: Simulate Test will cause fictitious test results to be generated in order to test communications
and demonstrate the unit. Do not forget to turn this off before deploying in a real application!
11.2.1.2.2 Suggested Communication Settings
Press the “Communications...” button to open the Communications dialogue. See figure 6.30
Select the Interface, Node Number and Baud Rate as shown, then press
“Use Defaults” to assign the
“base address”. This will define the start of the block of CAN message identifiers used by the software
(using a value compatible with the J1939 standard).