27 of 33
8.1 Technical data
8.1.1 Analog inputs
Thermocouples
Designation
Standard
Measuring range
Measuring accuracy
Fe-CuNi L
DIN 43710
(-200 to 900) °C
± 0.1 %
Fe-CuNi J
DIN EN 60584
(-210 to 1200) °C ± 0.1 % from -100 °C
Cu-CuNi U
DIN 43710
(-200 to 600) °C
± 0.1 % from -100 °C
Cu-CuNi T
DIN EN 60584
(-270 to 400) °C
± 0.1 % from -150 °C
NiCr-Ni K
DIN EN 60584
(-270 to 1372) °C ± 0.1 % from -80 °C
NiCr-CuNi E
DIN EN 60584
(-270 to 1000) °C ± 0.1 % from -80 °C
NiCrSi-NiSi N
DIN EN 60584
(-270 to 1300) °C ± 0.1 % from -80 °C
Pt10Rh-Pt S
DIN EN 60584
(-50 to 1768) °C
± 0.15 % from 20 °C
Pt13Rh-Pt R
DIN EN 60584
(-50 to 1768) °C
± 0.15 % from 50 °C
Pt30Rh-Pt6Rh B DIN EN 60584
(0 to 1820) °C
± 0.15 % from 400 °C
W5Re-W26Re C
(0 to 2320) °C
± 0.15 %
W3Re-W25Re D
(0 to 2495) °C
± 0.25 %
W3Re-W26Re
(0 to 2400) °C
± 0.15 %
Chromel-Copel
(-200 to 800) °C
± 0.1 % from -80 °C
Chromel-Alumel
(-200 to 1372) °C ± 0.1 % from -80 °C
PLII (Platinel II)
(0 to 1395) °C
± 0.15 %
MoRe5-MoRe41
(0 to 2000) °C
± 0.2 %
Cold junction
Pt1000 internal
Cold junction
accuracy
± 1 °C