6
External input/output functions MG-S series operation manual
-
66
-
6-5-3
Input command composition 2
Composed of 15 characters including a terminator (CR=0x0D/LF=0x0A)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
C1
C2
,
C3
C3
C3
C3
C3
C3
C3
C3
C3
C3
CR
LF
(1) ‘C3’ is maximum ten-digit (including the po/-, comma and point) numeric data.
Example) Upper limit input 120.00g: “LA,120.00”
Preset tare input 100.00g: “PT,100.00”
Interval time input 12:34:56: “IA,12,34,56” (marked off by commas)
(2) Make sure not input the measuring unit (g, c, etc.).
(3) Input the command when Weighing mode, Percentage mode, Counting mode or Multiplied by
Coefficient mode is operating.
If it is input while the other mode operation, the balance output an abnormal response.
(4) If the input value is invalid, the balance output an abnormal response.
(5) When the normal response, the preset tare value is input in <321 PRESET 1> and the balance
operates Preset tare.
(6) If the input value is “0” at Preset tare setting value command, the preset tare operation is
canceled.
6-5-3 (1)
Comparator setting command
C1
C2
Code
(
C1
)
Code
(
C2
)
Description
C3
Response
A00/Exx
format
ACK/NAK
format
L
A
0x4C
0x41 Upper limit
value setting
Numeric
value setting
A00:
Normal response
E01:
Abnormal response
ACK:
Normal response
NAK:
Abnormal response
L
B
0x4C
0x42 Lower limit
value setting
Numeric
value setting
L
C
0x4C
0x43 Reference
value setting
Numeric
value setting
6-5-3 (2)
Preset tare value setting command
This command is invalid for verified balance.
C1
C2
Code
(
C1
)
Code
(
C2
)
Description
C3
Response
A00/Exx
format
A00/Exx
format
P
T
0x50
0x54 Preset tare
value setting
Numeric
value setting
A00:
Normal response
E01:
Abnormal response
ACK:
Normal response
NAK:
Abnormal response
(1) When the normal response, the preset tare value is input in <321 PRESET 1> and the balance
operates Preset tare.
(2) If the input value is “0” at Preset tare setting value command, the preset tare operation is canceled.
6-5-3 (3)
Interval (output) time setting command
C1
C2
Code
(
C1
)
Code
(
C2
)
Description
C3
Response
A00/Exx
format
A00/Exx
format
I
A
0x49
0x41
Interval
(output) time
setting
Numeric
value setting
A00:
Normal response
E01:
Abnormal response
ACK:
Normal response
NAK:
Abnormal response
(1) If the input value is “0” at interval time setting command, the interval output operation is canceled.
(2) The initial setting value of interval time is “0”.
Summary of Contents for MG-S Series
Page 2: ......
Page 10: ......
Page 67: ...MG S series operation manual 6 External input output functions 57 ...
Page 68: ...6 External input output functions MG S series operation manual 58 ...
Page 84: ...8 Controlling and adjustment functions MG S series operation manual 74 ...
Page 109: ...MG S series operation manual Appendix 99 Appendix 7 Windshield assembly instructions ...
Page 110: ...Appendix MG S series operation manual 100 ...
Page 111: ...MG S series operation manual Appendix 101 ...
Page 113: ......
Page 114: ......