MG-S series operation manual 6
External input/output functions
-
65
-
6-5-2
Input command composition 1
Composed of four characters including a terminator (CR=0x0D/LF=0x0A).
1
2
3
4
C1
C2
CR
LF
6-5-2 (1)
Zero-point adjustment/Tare/Output control setting command
Please take care not to take alphabetical “O” for Arabic number “0”.
C1
C2
Code
(
C1
)
Code
(
C2
)
Description
Response
A00/Exx
format
ACK/NAK
format
T
(SP)
0x54
0x20
Tare subtraction
A00:
Normal
response
E01:
Abnormal
response
ACK:
Normal
response
NAK:
Abnormal
response
Z
(SP)
0x5a
0x20
Zero-point adjustment
O
0
0x4f
0x30
Stop output.
O
1
0x4f
0x31
Continuous output at all times
O
2
0x4f
0x32
Continuous output at stable times
(Output stop at unstable times)
O
3
0x4f
0x33
Push down [Output] key for one-time
instant output.
O
4
0x4f
0x34
Auto output
O
5
0x4f
0x35
One-time output at stable times
(Output stop at unstable times)
O
6
0x4f
0x36
One-time output at stable times
(Continuous output at unstable times)
O
7
0x4f
0x37
Push down [Output] key for one-time
output at stable times.
O
8
0x4f
0x38
One-time instant output
O
9
0x4f
0x39
One-time output after stability is obtained
O
A
0x4f
0x41
Interval function (Output once each time
the output time has elapsed)
O
B
0x4f
0x42
Interval function (Output once during
stabilization, each time the output time has
elapsed)
(1) Commands O8 and O9 are used to request data from the balance.
(2) Once the O0 to O7 commands are executed, that state is maintained. However, the status is
reset to the setting menu when the balance is turned on again.
(3) When the OA or OB command is input, the interval function starts, and when input again, the
interval function ends.
(4) After the O8 or O9 command is executed, it returns to “O0.”
6-5-2 (2)
Date output request and time output request
C1
C2
Code
(
C1
)
Code
(
C2
)
Description
Response
D
D
0x44
0x44
Date output request
Date data
D
T
0x44
0x54
Time output request
Time data
Summary of Contents for MG-S Series
Page 2: ......
Page 10: ......
Page 67: ...MG S series operation manual 6 External input output functions 57 ...
Page 68: ...6 External input output functions MG S series operation manual 58 ...
Page 84: ...8 Controlling and adjustment functions MG S series operation manual 74 ...
Page 109: ...MG S series operation manual Appendix 99 Appendix 7 Windshield assembly instructions ...
Page 110: ...Appendix MG S series operation manual 100 ...
Page 111: ...MG S series operation manual Appendix 101 ...
Page 113: ......
Page 114: ......