background image

1

 

Prior to use                                               MG-S series operation manual 

4

 

1-2-3

 

Precautions related to a specimen 

 

Static electricity 

 

In general, synthetic resin- and glass-made specimens are high in electric 
insulation, and so easily charged electrically. Weighing an electrically charged 
specimen makes the indication value unstable, reducing the reproducibility of 
the test result. Therefore, neutralize an electrically charged specimen before 
measurement. 

Magnetism 

 

Specimens affected by magnetism show different weight in a different position 
of the weighing pan, reducing the reproducibility. 
When weighing a magnetized specimen, either eliminate the magnetism from 
the specimen or place a setting plate on the weighing pan to distance the 
specimen from the weighing mechanism of the balance so that the mechanism 
may not be affected by the magnetism. 

Moisture 
absorption/ 
Evaporation 

 

Measuring a moist or evaporating (vaporizing) specimen increases or 
decreases the indication value of the balance continuously. When this is the 
case, put the specimen in a container equipped with a small mouth and closely 
seal the mouth before measurement. 

Specimen 
temperature 

 

Difference in temperature between the specimen and the windshield interior 
generates convection flow within the windshield, causing a measurement error. 
When the specimen temperature is excessively high or low, allow the specimen 
temperature to stabilize at the room temperature before measurement. Also, to 
prevent the convection flow from arising within the windshield, make the 
windshield interior temperature equal to the room temperature before 
measurement. 

 

 

Measurer’s body temperature also affects measurement result. Handle a 
specimen with tweezers instead of directly holding it with fingers and refrain 
from putting your hands directly in the windshield during measuring operation. 

 

1-2-4

 

Precautions related to the main unit of a balance 

 

Operating 
precautions 

 

A dust cover, if equipped, for the balance may possibly make the weight 
indication unstable due to static electricity charged on the cover at a low 
humidity. When this is the case, wipe the cover with wet cloth or use antistatic 
agent or use the balance with the cover removed. 

 

 

For more stable measurement, it is recommended to energize the balance for 
longer than 30 minutes and load the balance a few times with a weight 
equivalent to the weighing capacity before measurement. 

Calibration 

 

Calibrate the balance periodically with an external calibration weight. For the 
sake of precise calibration, use an external calibration weight weighing nearly 
equal to the weighing capacity of the balance. 

 

 

Energize the balance for longer than 30 minutes and load the balance a few 
times with a weight equivalent to the weighing capacity before calibration. 

 

 

Calibration is also needed in the following cases: 
When using the balance for the first time, 
When using the balance after a long period of non-use, 
When changing a place of installation, and 
When there was a large change in temperature, humidity or atmospheric 
pressure. 

Maintenance 

 

Attachment of dirt such as powder or liquid to the weighing pan or pan base will 
cause measurement error or unstable weight indication. For that reason, 
frequent cleaning of the balance is required. In cleaning the balance, take care 
for the dust or liquid not to enter into the balance. 

 

Summary of Contents for MG-S Series

Page 1: ...alance MG S Series Operation Manual S t a r M i c r o n i c s IMPORTANT To ensure safe and proper use of the balance please read this manual carefully After reading this manual store it in a safe place near the balance so you can review it as needed ...

Page 2: ......

Page 3: ...ng or irregularly bound will be exchanged Please inform the store where you purchased the product Trouble related to the product or system will be dealt with in accordance with the individual maintenance contract Please note however that we will not take responsibility for consequential trouble such as discontinuation of operation caused by the product trouble is the registered trademark of Star M...

Page 4: ...it is safe to think that what is not written in the operation manual cannot be performed unless the operation manual positively writes it is possible When performing installation operation maintenance or inspection of this product not only observe what is written or indicated in this document or on the product surface but also pay adequate consideration to safety measures The copyright of this doc...

Page 5: ...te the weighing data or for accurate weighing and appropriate usage of the equipment Used for reference information on operation Used for Prohibition items Used for Mandatory items requiring positive action Used for prohibition items to avoid Electrical shock This symbol indicates the operation specification related to NTEP approval and verification This product The product The balance Refers to t...

Page 6: ...ability and the response speed of the balance 5 User information setting Describes about setting items related to the upper and lower limits and preset tare weight 6 External input output functions Describes about setting items related to the specifications and conditions in regard to the external communication 7 Functions related to the lock Describes about setting items related to change prohibi...

Page 7: ...haracter font 13 2 Basic usage 14 2 1 Turning on off the power and checking for the operation 14 2 2 Zero point adjustment 15 2 2 1 Zero point adjustment range 15 2 3 Weighing a sample placed on a container tare 15 2 4 Weighing the additional sample 17 2 5 Basic operation 17 2 5 1 Hierarchy of a setting menu 17 2 5 2 Operation of the setting menu 18 2 5 3 Numeric value input 19 2 5 4 F key switchi...

Page 8: ...Preset tare 52 5 2 1 Preset tare setting 52 5 2 2 Inputting of a preset tare weight value 52 5 2 2 1 Actual value setting method 53 5 2 2 2 Numeric value setting method 53 5 2 2 3 Exiting the preset tare mode 53 5 3 Setting of the discrimination value of the comparator function 54 5 3 1 Actual value setting method 55 5 3 2 Numeric value setting method 55 6 External input output functions 56 6 1 Hi...

Page 9: ...5 1 Balance ID setting 80 8 5 2 Password control 80 8 5 2 1 Administrator password registration 81 8 5 2 2 User password registration 81 8 5 3 Outputting of the span calibration test result 82 8 5 4 Date indication format 82 8 5 5 Date setting 82 8 5 6 Time setting 83 8 5 7 Direct start setting 84 8 5 8 Initialize 84 9 Troubleshooting 85 9 1 Error message 85 10 How to maintain 88 10 1 Simple Metho...

Page 10: ......

Page 11: ...roduct was purchased Do not move the product with a sample to be weighed set on the balance That may cause the sample to fall from the weighing pan leading to a bodily injury or destruction of the sample Do not route the AC cord across passages The cord could be tripped on by a passerby and the balance could fall down and break or injure someone Do not use the product on an unstable table or a pla...

Page 12: ...sure to adjust calibrate the balance Check for an error periodically Use environment and chronological change cause an error in measured value leading to an inaccurate measurement Unplug the AC adapter from the receptacle when the balance is not going to be used for a long period of time Unplug the balance from the receptacle to save energy and prevent degradation Always adjust the level of the ba...

Page 13: ...enerate abrupt temperature change and resultantly cause unstable weight indication and therefore should be avoided Gravity The latitude and altitude of a measuring location differentiate the gravity that affects a specimen giving a different weight indication to the same specimen Electromagnetic wave At a location where a strong electromagnetic wave generating object is in the proximity of a balan...

Page 14: ...he room temperature before measurement Measurer s body temperature also affects measurement result Handle a specimen with tweezers instead of directly holding it with fingers and refrain from putting your hands directly in the windshield during measuring operation 1 2 4 Precautions related to the main unit of a balance Operating precautions A dust cover if equipped for the balance may possibly mak...

Page 15: ...y mounted on the verified balance Round pan type MG S322 Main unit Round 1 Round pan 1 Pan base Round 1 AC adapter 1 Operation manual 1 Sealing kit Wire sealing kit and tamper proof stickers Windshield Assembly type 1 Refer to Appendix 7 Windshield assembly instructions Square pan type MG S1501 MG S8200 Main unit Square 1 Square pan 1 Pan base Square 1 Pan base screw 1 AC adapter 1 Operation manua...

Page 16: ... and function of each section Round pan type MG S322 Square pan type MG S1501 MG S8200 1 Windshield 2 Weighing pan 3 Level 4 Adjuster 5 Display 6 Battery case 7 AC adapter jack 8 USB connector Type B 9 RS 232C connector D sub 9 pin male 10 Bluetooth module ...

Page 17: ... the balance Round pan type MG S322 1 Attach the Pan base Slider to check that in the Unlock side then attach to the balance 2 Move the Slider to Lock side 3 Mount the weighing pan 4 Assemble the windshield Refer to Appendix 7 Windshield assembly instructions to assemble the windshield 5 Connect the AC adapter ...

Page 18: ...embling the balance Square pan type MG S1501 MG S8200 1 Attach the Pan base 1 Direct FRONT to the display side 2 Attach to the balance then turn the Pan base screw to fix 2 Tighten the Pan base screw firmly 3 Mount the weighing pan 4 Connect the AC adapter ...

Page 19: ...gure on the left to loosen them 2 Level the balance 1 While watching the level turn the adjusters provided on the bottom to level the main unit 2 Bring the bubble enters in the center circle as shown in the figure on the left 3 When having leveled the main unit slightly push the four corners of the balance to make sure that there is no rattle Turn the adjusters as shown below depending on the posi...

Page 20: ... Use for selecting the mode function and item Use for moving down to the menu item selections or use for decrementing the numeric value 8 F3 F key Use for selecting the mode function and item Use for moving to the upper menu layer or use for selecting the digit to change 9 F4 F key Use for selecting the mode function and item Use for moving to the lower menu layer or use for selecting the digit to...

Page 21: ...o the setting menu 2 Tare Input a decimal point in Specific Gravity mode 3 Zero Use for changing polarity 4 F1 F key Use for incrementing the numeric values 0 1 2 9 0 5 F2 F key Use for decrementing the numeric values 0 9 8 1 0 6 F3 F key Use for selecting the digit to change 7 F4 F key Use for entering the value The F keys on which or are displayed above are available ...

Page 22: ...ndicates gross weight 8 Net weight Indicates that the tare weight is being subtracted 9 Preset tared weight Indicates that the preset tare weight is being subtracted 10 Gram Indicates the gram unit 11 milligram Indicates the milligram unit 12 16 segment message 16 segment unit Displays various messages Indicates the various units 13 Operation of the F key Displayed when the F1 F4 keys are effectiv...

Page 23: ...F G H I J K L M N O P Q R S T U V W X Y Z c comma point 1 2 3 4 5 6 7 8 9 0 space minus hyphen 16 segment A B C D E F G H I J K L M N O P Q R S T U V W X Y Z b c d g l m n o t w 1 2 3 4 5 6 7 8 9 0 asterisk slash left arrow right arrow space plus minus hyphen comma point percent Degree Celsius ...

Page 24: ...he self check the LCD display automatically changes Completion of the self check is followed by the weight mode Do not push any key during the self check 2 Balance operation check Press the weighing pan lightly to check if the indication changes 3 Turn off the power for the balance Push and hold On Off key About 2 seconds 1 Pushing and holding On Off key obtains the standby status from any operati...

Page 25: ...T STABLE For verified balance 17 WT STABLE is not indicated and the balance always wait stability during the zero point adjustment 2 2 1 Zero point adjustment range There is a Zero point adjustment range limit in this product When the weighing load gross exceeds the upper or lower limit Zero point adjustment cannot be executed Model Lower limit g Upper limit g MG S322 4 80 4 80 MG S1501 22 5 22 5 ...

Page 26: ...arrows the weighing range as much as the amount of the tare weight mass tare weight Weighable range weighing capacity tare weight 2 Tare weight can be output at 3 Check the tare weight by pushing Output key Check External input output functions to refer the output setting 3 Stability waiting during the tare can be set using the Setting menu 17 WT STABLE The setting of 17 WT STABLE is not changeabl...

Page 27: ...cation changes to zero and the symbol appears 3 Place an additional sample to be weighed The mass of the added sample alone is indicated 2 5 Basic operation Shortcuts for various modes functions can be assigned to F keys Please refer to 8 2 Shortcut setting for accessing various measuring modes and 8 3 Free key setting 2 5 1 Hierarchy of a setting menu The setting menu of this product is divided i...

Page 28: ...eration manual 18 2 5 2 Operation of the setting menu To perform settings for various functions from the state of weighing chiefly execute the following procedure Go to the menu item to set Select the setting value and execute fix ...

Page 29: ... e g When inputting 5 4321 1 Input Push Zero key to change the polarity to 2 Input 1 The digit for inputting is blinking Push F1 F2 key to increment decrement the digit to 1 Push F3 key to input the next digit 3 Input 2 3 4 5 Input 2 3 4 5 by the procedure above 4 Input Push Tare key to input on the immediately right of the blinking digit 5 Fix the input value Push F4 key to fix the input value 5 ...

Page 30: ...keys Weighing mode Counting mode Percentage mode Multiplied by Coefficient mode F1 F3 Free key Counting mode menu Percentage mode menu Multiplied by coefficient mode menu F4 F6 Free key Adding execution Sum total Adding execution Sum total Adding execution Sum total Comparator setting menu Comparator setting menu Comparator setting menu Comparator setting menu Measuring mode switching Measuring mo...

Page 31: ...For verified balance Gray shaded items are not indicated 17 WT STABLE is not indicated and fixed to be ON 141 ACTIVATE is not indicated and fixed to be OFF grain is not selectable on MG S1501 and MG S8200 carat is not selectable on MG S8200 Initial setting value of 12 UNIT is g gram on MG S322 MG S1501 and non verified MG 8200 LB pound on verified MG S8200 Initial setting value in a verified balan...

Page 32: ...y Please refer to 8 3 Free key setting 1 Select the weighing mode Push Menu key then push F1 F4 keys to go to 11 MODE Push F4 key to change the setting value Push F1 F2 key to select WEIG Weighing mode Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the weighing mode 3 Execute tare subtraction Place the container on the weighing pan if necessary Push Tare key Tare subtraction ...

Page 33: ...s to go to 11 MODE Push F4 key to change the setting value Push F1 F2 key to select COUN Counting mode Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the Counting mode 3 2 2 1 Actual value setting method Place the specified number of samples on the balance to record the average unit weight internally For verified balance Minimum sample size in pieces MSS is 10 PC On 5 in Step...

Page 34: ...that you are intended to count Push F4 key to fix the updated unit weight 6 Put samples in place to count them Place the samples Count result is displayed 1 When on VAR is selected in step 2 select the specified number of the sample among 1 to 999 by operating F1 F2 keys For verified balance 1 to 9 PC cannot be selected on on VAR 2 When simple SCS is operating if the weight of the samples is less ...

Page 35: ...e previous data When there is no data record this step is skipped Push F3 F4 key to select NO Change YES Not Change When YES is selected go to step 4 2 Select the unit weight value input mode Push F1 F2 key to select PCSWGT Unit weight value input Push F4 key to fix 3 Input the unit weight Numeric value input Refer to 2 5 3 Numeric value input Input the unit weight Push F4 key to fix 4 Put samples...

Page 36: ...s related to the operation MG S series operation manual 26 3 2 2 3 Switching the display at Counting mode 1 Push F1 F4 keys to switch the display For non ADD and TOTL can be used when 14 ADDITION is activated For ...

Page 37: ... of reference weight 1 Lower weight limit Reference weight Lower weight limit X 10 0 1 Lower weight limit X 10 Reference weight Lower weight limit X 100 0 01 Lower weight limit X 100 Reference weight 1 Select the percentage mode Push Menu key then push F1 F4 keys to go to 11 MODE Push F4 key to change the setting value Push F1 F2 key to select PCNT Percentage mode Push F4 key to fix 2 Exit the set...

Page 38: ...e case of onW Place the reference weight on the balance Push F4 key to record In the case of NUM Numeric value input Refer to 2 5 3 Numeric value input Input the reference value Push F4 key to fix 6 Weigh the samples The ratio of the weight of the sample to the reference weight is indicated in percent ...

Page 39: ...s operation manual 3 Functions related to the operation 29 3 3 1 Switching the display at percentage mode 1 Push F1 F4 keys to switch the display For non ADD and TOTL can be used when 14 ADDITION is activated For ...

Page 40: ...d by Coefficient mode Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the Multiplied by Coefficient mode 3 Select whether or not employ the previous recorded coefficient Push F3 F4 key to select whether or not employ the previous data When there is no data record this step is skipped Push F3 F4 key to select NO Change YES Not Change When YES is selected go to step 6 4 Set the ...

Page 41: ... operation manual 3 Functions related to the operation 31 3 4 1 Switching the display at Multiplied by Coefficient 1 Push F1 F4 keys to switch the display ADD and TOTL can be used when the 14 ADDITION is activated ...

Page 42: ... liquid 3 Measure the sample weight in the air 4 Compensate the buoyancy acting on the net basket 5 Measure the sample weight in the water liquid 6 The specific gravity of the sample is displayed 1 Select the specific gravity mode Push Menu key then push F1 F4 keys to go to 11 MODE Push F4 key to change the setting value Push F1 F2 key to select SPGR specific gravity mode Push F4 key to fix 2 Exit...

Page 43: ...e net basket Remove the sample on the net basket and push Tare key to tare then sink the net basket into the water liquid Push Tare key to compensate the buoyancy acting on the net basket 7 Measure the sample weight in the water liquid Put the sample on the net basket in the water liquid then push F4 key to record The specific gravity of the sample for the 4 C water is automatically calculated and...

Page 44: ...put function The setting of 17 WT STABLE The output condition ON Once at stable after Output key is pushed OFF Once immediately after Output key is pushed 1 Select the statistics mode Push Menu key then push F1 F4 keys to go to 11 MODE Push F4 key to change the setting value Push F4 key to select STAT Statistics mode Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the statisti...

Page 45: ... required number of data items are collected 6 Display the statistical operation result Push F2 key DATA The display changes to the statistical operation display Push F1 F2 key to switch to another calculated item Sum total Maximum value Minimum value Range Max Min Average value Standard deviation Coefficient of variation 3 6 1 Switching the display at Statistics mode 1 Push F1 F4 keys to switch t...

Page 46: ...owing Output once after the indication is held except when the HOLD is pushed step 5 b Output once after the Output key is pushed during the indication is held 2 Only g can be used 1 Select the animal mode Push Menu key then push F1 F4 keys to go to 11 MODE Push F4 key to change the setting menu Push F1 F2 key to select ANIM Animal mode Push F4 key to fix 2 Exit the setting menu Push Menu key to s...

Page 47: ...set range the weighing indication is held and indication appears b When the animal is too wild weigh the animal using manual HOLD key Push F4 NEXT key to display the HOLD menu on F1 key Place the animal on the weighing pan Push F1 HOLD key then the weighing indication is held and indication appears 6 Remove the animal Remove the animal then automatically tare subtracted ...

Page 48: ...key is pushed OFF Once immediately after Output key is pushed 1 Select the formulation mode Push Menu key then push F1 F4 keys to go to 11 MODE Push F4 key to change the setting value Push F1 F2 key to select FORM Formulation mode Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the Formulation mode 3 Choose whether or not clear all the data Push F3 F4 key to select whether or ...

Page 49: ...Put the sample on the tare and push Output key to store the sample weight Repeat the step 6 for all the samples to be compounded When to set the tare individually for each sample repeat steps 4 6 3 8 1 Check the stored data of each component 1 Push F1 F4 keys to check the weight of each component ...

Page 50: ...n Weighing mode Percentage mode and Counting mode 3 10 1 How to perform discrimination Set the lower and the upper limits Then whether the weight of a sample to be weighed is LOW lower than the lower limit OK appropriate or HIGH higher than the upper limit is indicated on the LCD with 16 segment messages 16 segment messages Discrimination Single point setting lower limit Single point setting upper...

Page 51: ...limit alone valid Push F4 key to fix 2 Select the discriminant condition Push F1 F4 keys to go to 132 CONDITION Push F4 key to change the setting value Push F1 F2 key to select FULL At all times STBL Only at stable times Push F4 key to fix 3 Select the discriminant range Push F1 F4 keys to go to 133 RANGE Push F4 key to change the setting value Push F1 F2 key to select 5 5 d or more 50 50 d or mor...

Page 52: ...keys to go to 142 OPERATION Push F4 key to change the setting value Push F1 F2 key to select TOTAL Addition accumulated NET Net addition Push F4 key to fix 3 Select the adding direction Push F1 F4 keys to go to 143 DIRECTION Push F4 key to change the setting value Push F1 F2 key to select PLUS Plus side addition MINUS Minus side addition Push F4 key to fix 4 Set the Free key Set the following func...

Page 53: ...dication to zero and then place the next sample to be weighed After appears push F1 ADD key The weighed value is stored and SUM TOTAL is indicated for a few seconds Repeat this operation to perform addition In the case of the net addition Add a sample to be weighed The indication automatically returns to zero Then add the next sample to be weighed After appears push F1 ADD key After indicating SUM...

Page 54: ... After appears push F1 ADD key The weighed value is stored and SUM TOTAL is indicated for a few seconds Repeat this operation to perform addition In the case of the net addition Remove the sample Remove the sample to be weighed After appears push F1 ADD key After indicating SUM TOTAL and the accumulated value for a few seconds the balance returns to the weight indication followed by the automatic ...

Page 55: ...ted when the weighing indication is over the zero point adjustment range before tare subtraction and when the net indication is negative after tare subtraction 1 Select the tare subtraction reminder function Push Menu key then push F1 F4 keys to go to 15 T REMINDER Push F4 key to change the setting menu Push F1 F2 key to select OFF Invalid 1 Activate the Mode 1 2 Activate the Mode 2 Push F4 key 2 ...

Page 56: ...able the function OFF Disable ON Activate Push F4 key to fix 2 Exit the setting menu and operate with zero point adjustment reminder Push Menu key to exit the setting menu Put the samples on the weighing pan then remove it then the PUSH ZERO alert is displayed The alert disappears after Zero key is pushed and zero point adjustment is completed 3 14 Bar graph indication Set the indication non indic...

Page 57: ...ush F1 F2 key to select OFF Invalid ON Valid Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the measuring mode 3 16 Backlight setting Setting the backlight control 1 Select the backlight setting Push Menu key then push F1 F4 keys to go to 1A BACKLIGHT Push F4 key to change the setting value Push F1 F2 key to select Refer to the Set List Push F4 key to fix Set List OFF Invalid...

Page 58: ...s not work under the following conditions 1 Setting menu is being displayed 2 A sample is placed on the weighing pan and the display is not stable When is not displayed 3 18 Simple SCS Self Counting System method setting Simple SCS method is auxiliary function for Counting mode First put a set number of samples in place Next put up to two times the set number of additional samples in place The bal...

Page 59: ...tability discrimination width When the larger numeric value is set in this setting menu the laxer stability judgement is applied and the balance indicate Stable mark in more unstable conditions For verified balance 21 STABLE 2 4 are not available 1 Select the stability discrimination width Push Menu key then push F1 F4 keys to go to 21 STABLE Push F4 key to change the setting value Push F1 F2 key ...

Page 60: ... Slow 2 Exit the setting menu Push Menu key to shift to the measuring mode 4 4 Zero tracking Setting to the zero tracking function makes it possible to automatically correct the zero point fluctuation caused by the temperature fluctuation etc when 0 is indicated through which the 0 indication is maintained For verified balance 23 ZERO TRAC 1 2 4 are not available 1 Select the zero tracking Push Me...

Page 61: ...n setting 51 5 User information setting Describes about setting items related to the comparator function 5 1 Hierarchy of user information setting For verified balance gray shaded items are not indicated Initial setting value in a verified balance ...

Page 62: ...ist OFF Invalid 1 Setting 1 execute 2 Setting 2 execute 3 Setting 3 execute 4 Setting 4 execute 5 Setting 5 execute 2 Exit the setting menu Push Menu key to shift to the measuring mode Preset tared value is displayed with indication when preset tare value is available 5 2 2 Inputting of a preset tare weight value There are two ways of inputting a preset tare weight value described below Actual val...

Page 63: ...ft to the measuring mode is indicated 5 2 2 2 Numeric value setting method 1 Set a preset tare weight value Numeric value input Refer to 2 5 3 Numeric value input Push F4 key to select NUM Numeric value Input the preset tare value Push F4 key to fix The preset tare weight value is stored 2 Exit the setting menu Push Menu key to shift to the measuring mode is indicated 5 2 2 3 Exiting the preset ta...

Page 64: ...d by the upper and lower limit values that have been set for the reference numeric value For example Two point upper and lower limits setting Reference value 1000 0g Lower limit value 900 0 g Upper limit value 1200 0 g Discrimination method Reference value Lower limit value Upper limit value 1000 0 g 900 0 g 1200 0 g Absolute value 900 0 g 1200 0 g Relative value 1000 0 g 100 0 g 200 0 g 1 Select ...

Page 65: ... a sample to be weighed that is equivalent to the upper limit value Push F4 key to fix The upper limit value is recorded 5 3 2 Numeric value setting method 1 Set a lower limit value Numeric value input Refer to 2 5 3 Numeric value input Push F4 key to select NUM Numeric value Input the lower limit value Push F4 key to fix The lower limit value is saved 2 Set a reference value In the case of the re...

Page 66: ... The initial setting value of 411 421 431 ACTIVATE are OFF for non verified balance ON for verified balance 412 422 432 FORMAT are not indicated and fixed to be CBM for verified balance The initial setting value of 413 423 433 CONDITION are 7 for non verified balance 2 for verified balance 41A 42A 43A STATUS are Initially set to OFF for non verified balance Not indicated and fixed to be ON for ver...

Page 67: ...MG S series operation manual 6 External input output functions 57 ...

Page 68: ...6 External input output functions MG S series operation manual 58 ...

Page 69: ...xamples as a guide to connect the balance to external devices using the cable Sample connection with a PC AT compatible machine 2 D sub9P Connecter can execute tare subtraction or zero point adjustment from an external device by connecting a contact or a transistor switch between Pin 9 EXT TARE and Pin 5 GND When doing so allow at least 400 ms for connection ON time Maximum voltage 15 V when the b...

Page 70: ... digit numeric format CSP 6 digit format Consists of 14 characters including terminators CR 0xDH LF 0xAH 1 2 3 4 5 6 7 8 9 10 11 12 13 14 P1 D1 D2 D3 D4 D5 D6 D7 U1 U2 S1 S2 CR LF 7 digit numeric format CSP 7 digit format Consists of 15 characters including terminators CR 0xDH LF 0xAH 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 P1 D1 D2 D3 D4 D5 D6 D7 D8 U1 U2 S1 S2 CR LF 8 digit numeric format Consists o...

Page 71: ...mme O Z 0x4F 0x5A ounce L B 0x4C 0x42 pound O T 0x4F 0x54 troy ounce D W 0x44 0x57 pennyweight G R 0x47 0x52 grain T L 0x54 0x4C Hong Kong tael T L 0x54 0x4C Singapore Malaysia tael T L 0x54 0x4C Taiwan tael t o 0x74 0x6F tola M S 0x4D 0x53 mesghal B A 0x42 0x41 baht P C 0x50 0x43 parts counting SP 0x20 0x25 percentage weighing SP 0x20 0x23 Multiplied by Coefficient S1 one character Indicates the ...

Page 72: ...T1 T6 6 characters Represents the type of the data For non SP SP SP SP SP SP 0x20 0x20 0x20 0x20 0x20 0x20 Net weight 41A STATUS OFF N SP SP SP SP SP 0x4E 0x20 0x20 0x20 0x20 0x20 Net weight 41A STATUS ON P T SP SP SP SP 0x50 0x54 0x20 0x20 0x20 0x20 Preset tare weight T SP SP SP SP SP 0x54 0x20 0x20 0x20 0x20 0x20 Tare weight T O T A L SP 0x54 0x4F 0x54 0x41 0x4C 0x20 Total value Accumulated valu...

Page 73: ... 0x4F 0x54 troy ounce d w 0x64 0x77 pennyweight G R 0x47 0x52 grain t l 0x64 0x6C Hong Kong tael t l 0x64 0x6C Singapore Malaysia tael t l 0x64 0x6C Taiwan tael t o 0x74 0x6F tola M S 0x4D 0x53 mesghal B A 0x42 0x41 baht P C 0x50 0x43 parts counting SP 0x20 0x25 percentage weighing SP 0x20 0x23 Multiplied by Coefficient For SP g 0x20 0x67 gram SP c 0x20 0x63 carat o z 0x6F 0x7A ounce l b 0x6C 0x62...

Page 74: ...sful completion or if the command is invalid an error the balance will transmit an error response When the balance is in normal display mode it usually sends a response to a command within one second of receiving the command For the tare range a response is sent after the commands are completely processed 1 After you have sent an input command the balance return the response command approximately ...

Page 75: ...tput stop at unstable times O 6 0x4f 0x36 One time output at stable times Continuous output at unstable times O 7 0x4f 0x37 Push down Output key for one time output at stable times O 8 0x4f 0x38 One time instant output O 9 0x4f 0x39 One time output after stability is obtained O A 0x4f 0x41 Interval function Output once each time the output time has elapsed O B 0x4f 0x42 Interval function Output on...

Page 76: ...cription C3 Response A00 Exx format ACK NAK format L A 0x4C 0x41 Upper limit value setting Numeric value setting A00 Normal response E01 Abnormal response ACK Normal response NAK Abnormal response L B 0x4C 0x42 Lower limit value setting Numeric value setting L C 0x4C 0x43 Reference value setting Numeric value setting 6 5 3 2 Preset tare value setting command This command is invalid for verified ba...

Page 77: ...b9P Connecter can execute tare subtraction or zero point adjustment from an external device by connecting a contact or a transistor switch between the pin for externally setting a tare range Pin 9 and the signal ground pin Pin 5 When doing so allow at least 400 ms for connection ON time Maximum voltage 15 V when the balance is turned OFF sink current 20 mA when it is turned ON 1 While external con...

Page 78: ...n setting of your PC is required Please refer to Appendix 5 USB communication and bus power input 1 Select the RS 232C communication operation Push Menu key then push F1 F4 keys to go to 411 ACTIVATE Push F4 key to change the setting value Push F1 F2 key to select OFF Stop ON Operation Push F4 key to fix Select the USB communication operation Push Menu key then push F1 F4 keys to go to 421 ACTIVAT...

Page 79: ... than or equal to zero by unloading zero point adjustment or tare subtraction 5 One time output at stable times Output stop at unstable times 6 One time output at stable times Continuous output at unstable times 7 Push down Output key for one time output at stable times Select the comparator output Set list 0 As per the output setting 1 Output when discrimination result is OK or absent Select the ...

Page 80: ...ize pairing with external Bluetooth device 1 Select Bluetooth initialization Push Menu key then push F1 F4 keys to go to 44 BLUETOOTH BLE then push F4 key to go to 441 INIT 2 Initialize Bluetooth pairing Push F4 key to execute initialization of Bluetooth pairing 3 Exit the setting menu Push Menu key to shift to the measuring mode ...

Page 81: ...archy of functions related to the lock Initial setting value 7 2 Total lock release All locks that have been set can be released 1 Select the total lock release Push Menu key then push F1 F4 keys to go to 51 ALL UNLOCK Push F4 key Push F1 F2 key to select YES Execute NO NO execute Unlock all the settings 2 Exit the setting menu Push Menu key to shift to the measuring mode ...

Page 82: ...t to the measuring mode 7 4 Menu lock function Various setting menus can be locked 1 Select the menu lock function Push Menu key then push F1 F4 keys to go to 53 MENU LOCK Push F4 key to change Push F1 F2 key to select Refer to Set List Set list 531 OPERATION Function related to the operation 1 APPLICATIONS 532 PERFORM Function related to the performance 2 PERFORMANCE 533 USER User information set...

Page 83: ...s 73 8 Controlling and adjustment functions Perform setting of the Balance ID Scale ID the span calibration test and the date and time 8 1 Hierarchy of controlling and adjustment functions For verified balance gray shaded items are not available Initial setting value ...

Page 84: ...8 Controlling and adjustment functions MG S series operation manual 74 ...

Page 85: ...o 611 F1 KEY Push F4 key to change Push F1 F2 key to select 611 F1 KEY F1 above F1 key 612 F2 KEY F2 above F2 key 613 F3 KEY F3 above F3 key Push F4 key to fix 2 Select the measuring modes Push F4 key to change the setting value Push F1 F2 key to select Refer to Set List Push F4 key to fix Set list WEIG Weighing mode COUN Counting mode PCNT Percentage mode MULT Multiplied by Coefficient mode SPGR ...

Page 86: ...y setting menu list 621 F1 KEY F1 622 F2 KEY F2 623 F3 KEY F3 624 F4 KEY F4 625 F5 KEY F5 626 F6 KEY F6 2 Select the function to assign to the Free key Push F4 key to change the setting value Push F1 F2 key to select Refer to Set List Push F4 key to fix Set list NONE OFF B G Gross Net Date DAte indication TIME Time indication TARE Tare value indication HIGH Upper limit value LOW Lower limit value ...

Page 87: ... 2 The span calibration significantly affects the weighing accuracy Please read this procedure carefully before getting to the calibration 8 4 1 1 Span calibration with external weight 1 This mode is not available for verified balance 2 SELECT MIN 1 shall be selected at span calibration during the verification 1 Select the span calibration with external weight Push Menu key then push F1 F4 keys to...

Page 88: ...ring mode 1 List of weights used for the span calibration by model Unit g Model MG S322 MG S1501 MG S8200 Selectable weight on the menu 320 1500 8200 300 1000 8200 200 1000 5000 100 500 5000 50 200 2000 5 20 200 set 1 to 320 1 to 1500 1 to 8200 2 The span calibration by the use of a weight less than the weighing capacity may possibly indicate UC on the display When this is the case the weighing ac...

Page 89: ...anges to the order of on 0 blinking of on 0 On completion of the zero point adjustment the display automatically changes to on FS 3 The span test starts Place the weight in the center of the weighing pan Display changes to the order of on FS blinking of on FS Start of the span test On completion of the span test the display automatically changes to INSTRUMENTAL ER and the instrumental error of the...

Page 90: ...sh F1 F2 key Repeat the input by the procedure above Push F4 key to fix the balance ID and shift to 641 SCALE ID 3 Exit the setting menu Push Menu key to shift to the measuring mode 8 5 2 Password control Enable disable the password protection 1 Refer to 8 5 2 1 Administrator password registration and 8 5 2 2 User password registration for password registration changing 2 Refer to Appendix 6 Balan...

Page 91: ...o set the password Four digits of 0 9 can be selected Push F4 key to fix 3 Exit the setting menu Push Menu key to shift to the measuring mode 8 5 2 2 User password registration Administrator can register the user password for each user operator 1 Refer to Appendix 6 Balance operation with password control function for setting each user s authority 2 User password can be set only for User 1 and Use...

Page 92: ...ey to shift to the measuring mode 8 5 4 Date indication format Date indication format can be selected 1 Select the Date indication format Push Menu key then push F1 F4 keys to go to 646 DATE DISP Push F4 key to change the setting value Push F1 F2 key to select Y M D Year Month Day D M Y Day Month Year M D Y Month Day Year Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the mea...

Page 93: ...ime setting 1 Select the time setting Numeric value input Refer to 2 5 3 Numeric value input Push Menu key then push F1 F4 keys to go to 648 TIME SETTING Push F4 key to change the setting menu The digit for inputting is blinking Input the time Push F4 key to fix the time setting 2 Exit the setting menu Push Menu key to shift to the measuring mode 3 Indication of the time Push F3 TIME key The time ...

Page 94: ...e direct start Push Menu key then push F1 F4 keys to go to 64C DIRECT ST Push F4 key to change the setting value Push F1 F2 key to select OFF Disable ON Enable Push F4 key to fix 2 Exit the setting menu Push Menu key to shift to the measuring mode 8 5 8 Initialize This function is to initialize the balance to the factory settings except span calibration Bluetooth pairing the date and time setting ...

Page 95: ...e and notify the store where you purchased the product ERR703 The operation key was pushed at the time of starting from the standby status If the error message is displayed nevertheless the operation key wasn t pushed there is something wrong with the hardware Do not push the operation key while the balance is in the process of starting from the standby status ERR705 Initial zero adjustment error ...

Page 96: ...th other object Check for any wind or vibration ERR739 Time out error of importing the sample weight in the actual value setting method at Preset tare setting Improper setting of the weighing pan or pan base is suspected Check for contact with other object Check for any wind or vibration ERR740 The setting value is out of the setting range at numeric value setting method or actual value setting me...

Page 97: ...other data ERR755 Time out error of importing the sample weight at Statistics Formulation mode Improper setting of the weighing pan or pan base is suspected Check for contact with other object Check for any wind or vibration ERR756 Weight of the sample is out of the importing range at Statistics Formulation mode 0g to maximum capacity Choose the sample of which weight is within the importing range...

Page 98: ...ndshield 1 Refer to Appendix 7 Windshield assembly instructions to remove the windshield 2 Remove the round pan 3 Move the slider to Unlock side 4 Remove the pan base 5 Maintenance method 1 Wipe dirt from the balance with dry and soft cloth 2 In the case of heavy soil dismount the weighing pan and or the pan base and clean them with a piece of cloth slightly wet with neutral detergent or solvent ...

Page 99: ...ype MG S1501 MG S8200 1 Remove the square pan 2 Remove the pan base 3 Maintenance method 1 Wipe dirt from the balance with dry and soft cloth 2 In the case of heavy soil dismount the weighing pan and or the pan base and clean them with a piece of cloth slightly wet with neutral detergent or solvent ...

Page 100: ...BAt 320090 mg X External MG S1501 1500 g 7500 ct 3 3 lb 52 oZ 48 oZt 23000 GN 960 dwt 400 mom 320 MSG 40 tlH 39 tlS 40 tlT 120 to 98 BAt 1500000 mg 0 1 g 1 ct 0 001 lb 0 01 oZ 0 01 oZt 10 GN 0 1 dwt 0 1 mom 0 1 MSG 0 01 tlH 0 01 tlS 0 01 tlT 0 01 to 0 01 BAt 100 mg 0 1 g 1 ct 0 001 lb 0 01 oZ 0 01 oZt 10 GN 0 1 dwt 0 1 mom 0 1 MSG 0 01 tlH 0 01 tlS 0 01 tlT 0 01 to 0 01 BAt 100 mg 1500 9 g 7509 ct...

Page 101: ...lance Item Description Weighing system Tuning fork vibration method Measuring mode Weighing Counting Percentage Multiplied by Coefficient Specific gravity solid Statistics Animal Formulation Function Function related to the operation Unit setting Comparator Adding Tare subtraction reminder Zero point adjustment reminder Stabilization wait setting Bar graph Backlight Auto power off Simple SCS Funct...

Page 102: ... MG S1501 MG S8200 0 1 g 1 g 10 g Power Dedicated AC adapter 100 240VAC 50 60Hz 4 AA Dry cell batteries USB bus power Connected with PC in which the driver is installed Ratings AC adapter jack Battery box 4 AA batteries USB bus power 4 6VDC 4 6VDC 5VDC 0 3A 0 3A 0 3A Maximum current consumption Dimensions of the weighing pan MG S322 MG S1501 MG S8200 φ118mm 160 x 180mm Weight of the balance NET Ap...

Page 103: ...MG S series operation manual Appendix 93 Appendix 2 Dimensional outline drawing MG S322 MG S1501 MG S8200 ...

Page 104: ...1 00000000E 00 carat 5 00000000E 00 pound 2 2046226E 03 ounce 3 5273961E 02 troy ounce 3 2150746E 02 grain 1 5432358E 01 pennyweight 6 4301493E 01 momme 2 6666667E 01 mesghal 2 16999761E 01 Hong Kong tael 2 6717251E 02 Singapore Malaysia tael 2 6455471E 02 Taiwan tael 2 6666667E 02 tola 8 5735324E 02 baht 6 59630607E 02 milligram 1 00000000E 03 ...

Page 105: ... with the positive and negative poles correctly inserted 3 Insert the battery case Insert the battery case until it clicks in place When the balance is battery operated is displayed It changes in accordance with the remaining battery capacity Mark Description The battery level is sufficient The battery level is low The batteries have run down Replace them with new ones Continuous battery runtime A...

Page 106: ...he Device Manager Window 1 1 How to open the Device Manager Window For Windows 7 Go to Start Menu Right click the Computer Properties Device Manager For Windows 10 Right click the Start button Go to Device Manager 2 Click the Port COM and LPT to open the thread and double click the STAR MG S COM to open the properties window 3 Go to the Power Management tab of the Star MG S COM properties window 4...

Page 107: ...4 key to go to Administrator login mode is indicated on the 7 segment display 3 Select the user to set the authority Select the user 0 Guest user 1 User 1 2 User 2 USER Shift to the User login mode 4 Input the administrator password Input the administrator password by pushing F1 F4 keys Each digit increments as 0 1 8 9 0 by pushing each F key First digit from the left F1 key Second digit from the ...

Page 108: ...and or accessing to setting menus Appendix 6 2 User guest login 1 Power on the balance and go to the User login mode Self check Push On Off key then the balance shifts to User login mode after start up operation is indicated on the 7 segment display 2 Select the user number Select the user operator number 0 Guest user 1 User 1 2 User 2 ADMIN Shift to the Administrator login mode 3 Input the user p...

Page 109: ...MG S series operation manual Appendix 99 Appendix 7 Windshield assembly instructions ...

Page 110: ...Appendix MG S series operation manual 100 ...

Page 111: ...MG S series operation manual Appendix 101 ...

Page 112: ...4 measuring mode 20 22 75 menu lock function 72 multiplied by coefficient mode 30 numeric value 11 19 numeric value setting method 27 52 54 operation keys 10 output conditions 69 parity bit 69 password 80 97 percentage mode 20 27 75 preset tare 52 66 range 35 reference value 40 54 response command 69 response speed 50 RS 232C 56 segment 12 setting menu 18 setting value 11 18 Simple SCS method 48 s...

Page 113: ......

Page 114: ......

Reviews: