T15e Tag Deployment & User Guide
60
Show/Hide Thresholds
You can choose to either show or hide the temperature thresholds on the main
screen by doing the following:
1.
Press the Select button
to access the menu.
2.
Navigate to the ‘Show/Hide Thresholds’ option using the arrow buttons
.
3.
Press the Select button
.
Performing an Audit – Manual Inspection
For VFC deployments, the CDC requirements include performing a manual
inspection of the device twice a day (12 hours apart) to verify that it is operating
normally. This activity is called an "Audit."
Pressing and holding the main button
for three to five seconds displays
‘Audit report has been recorded successfully’. This validates the inspection
in accordance with CDC requirements. The operation is also logged by the tag
and is sent to MobileView along with the current temperature, and the
Minimum and Maximum Temperature values recorded since the last Audit
.
The
Min and Max measurement values of the last audit are also renewed.