Chapter 4
Programming
125
SR860 DSP Lock-in Amplifier
FFT Screen Commands
The SR860 has a basic FFT function accessable from the front panel by selecting the appropriate [Screen
Layout] display (remotely, by setting
DLAY FFT
or
DLAY BARFFT
). A more complete discussion of FFT
mode can be found in Appendix B .
Note that while FFT configuration commands and queries may be sent at any time, the FFT data query
(
FCRY?
) only returns valid data while the display is showing the FFT.
FFTR(?) { ADC | MIXer | FILTer | i }
The
FFTR i
command sets the source for the FFT to input ADC (i=0), post mixer PSD
(i=1) or post time constant filter (i=2).
Example
FFTR FILT
Set the FFT source to post time constant filter
FFTR 2
FFTR?
Returns the FFT source i
FFTS(?) { i }
The
FFTS i
command sets the FFT vertical scale (dB/div) according to the table below.
i
dB/div
i
dB/div
-8
2 m
0
1.0
-7
5m
1
2.0
-6
10 m
2
5.0
-5
20 m
3
10
-4
50 m
4
20
-3
100 m
5
50
-2
200 m
6
100
-1
500 m
7
200
The argument i is limited to
−
20
≤ i ≤
20. Vertical scale changes preserve the amplitude
of the top edge of the graph.
Example
FFTS 4
Set the FFT vertical scale to 20 dB/div
FFTS?
Returns the FFT vertical scale i
FFTO(?) { x }
The
FFTO x
command sets the FFT vertical offset x (dB). This is the amplitude
associated with the top edge of the graph.
Example
FFTO -20.0
Set the FFT vertical offset to
−
20.0 dB
FFTO?
Returns the FFT vertical offset
Summary of Contents for SR860
Page 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Page 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Page 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Page 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Page 118: ......
Page 172: ......
Page 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Page 192: ......