108
Programming
Chapter 4
SR860 DSP Lock-in Amplifier
Example
HARMDUAL 2
Set the dual external harmonic detect to 2
HARMDUAL?
Returns the dual external harmonic number i
BLADESLOTS(?) {SLT6 | SLT30 | i }
The
BLADESLOTS i
command configures the lock-in for operation with an external
SR540 chopper. Before proper operation in
RSRC EXT
mode, the SR860 must be
configured for either 6-slot (i=0) or 30-slot (i=1) operation, based on the hardware
installed on the SR540 chopper. You can read more about operation in chop mode on
page 76.
Example
BLADESLOTS SLT30
Configure for 30-slot chopper blade
BLADESLOTS 1
BLADESLOTS?
Returns chopper blade configuration
BLADEPHASE(?) { p } { UDEG | MDEG | DEG | URAD | MRAD | RAD }
The
BLADEPHASE p
command sets the phase of the SR540 chopper blade. When
operating a single chopper, this has little effect since the SR860 will follow the chopper.
If multiple SR860/SR540 systems are being used in a single experiment,
BLADEPHASE
can be used to modify the relative phase of the choppers. Blade phase may be specified in
degrees (default), or millidegrees, microdegrees, radians, milliradians, or
microradians.You can read more about chopper blade phase on page 77.
Example
BLADEPHASE 43.21 DEG
Set the blade phase to 43.21 deg
BLADESLOTS 4.321E1
BLADEPHASE?
Returns chopper blade phase
SLVL(?) { v } { NV | UV | MV | V }
The
SLVL v
command sets the sine out amplitude to v. The value of v will be rounded to
3 digits or 1 nV, whichever is greater. The amplitude may be programmed from 1 nV to
2.0 V.
Example
SLVL 12.3 MV
Set the sine out amplitude to 12.34 mV
SLVL 1.23E-2
SLVL 0.0123
SLVL?
Returns the sine out amplitude in Volts
SOFF(?) { v } { NV | UV | MV | V }
The
SOFF v
command sets the sine out dc level to v. The value of v will be rounded to 3
digits or 0.1 mV, whichever is greater. The level may be programmed from
−5.00
V to
+5.00 V.
Summary of Contents for SR860
Page 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Page 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Page 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Page 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Page 118: ......
Page 172: ......
Page 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Page 192: ......