3.4 Commands
3 – 15
Amplified Error Monitor
EMON? [
i
]
Query the
P
×
ε
voltage, in volts.
i
is an optional parameter that causes streaming of (
P
×
ε
) data. If
i
is specified, then
i
measurements will be output at a rate of approx-
imately half a second per measurement. If
i
is specified as 0, then
measurements will be output indefinitely. The SOUT command can
be used to stop streaming.
EMON?
Example:
+00.105537
Output Monitor
OMON? [
i
]
Query the
Output
voltage, in volts.
OMON?
always reports the voltage generated at the front-panel OUT-
PUT BNC connector (regardless of the state of
AMAN
).
i
is an optional parameter that causes streaming of
Output
data.
If
i
is specified, then
i
measurements will be output at a rate of
approximately half a second per measurement. If
i
is specified as 0,
then measurements will be output indefinitely. The SOUT command
can be used to stop streaming.
OMON?
Example:
+01.106139
Output Streaming Records Format
RFMT(?)
{
z
}
Set (query) the output streaming record format
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
When
ON
, data are output on a single line with three comma delim-
iters. Since there are four monitor channels that can be streamed to
output, and any combination of the four may be streamed, the comma
delimiters allow unambiguous identification of channel data.
The record format is
SMON
,
MMON
,
EMON
,
OMON
RFMT ON
Example:
SMON? 3; MMON? 3; EMON? 3; OMON? 3
+00.099909,-00.00.01.106135
+00.099909,-00.00.01.106123
+00.099915,-00.00.01.106151
SIM960 Analog PID Controller