3 – 14
Remote Operation
Lower Output Limit
LLIM(?)
{
f
}
Set (query) the lower output limit
{
to
f
}
, in volts.
The lower limit can be set with 10 mV resolution. The output voltage
of the SIM960 will always be clamped to remain less negative than
the
LLIM
limit. See
ULIM
for more details.
The allowed range is
−
10
.
00
≤
LLIM
≤
ULIM
.
3.4.5
Monitor commands
Setpoint Input Monitor
SMON? [
i
]
Query the
Setpoint
input voltage to the error amplifier, in volts.
If
INPT
INT
is set, then
SMON?
monitors the value of the internally-
generated setpoint. If
INPT
EXT
, then
SMON?
monitors the voltage
applied at the front-panel Setpoint BNC input.
i
is an optional parameter that causes streaming of
Setpoint
data.
If
i
is specified, then
i
measurements will be output at a rate of
approximately two measurements per second. If
i
is specified as 0,
then measurements will be output indefinitely. The SOUT command
can be used to stop streaming.
SETP 1.2; SMON? 5
Example:
+01.004496
+01.066567
+01.128909
+01.191273
+01.200073
Measure Input Monitor
MMON? [
i
]
Query the
Measure
input voltage to the error amplifier, in volts.
MMON?
always reports the voltage applied at the front-panel Mea-
sure BNC input.
i
is an optional parameter that causes streaming of
Measure
data.
If
i
is specified, then
i
measurements will be output at a rate of
approximately half a second per measurement. If
i
is specified as 0,
then measurements will be output indefinitely. The SOUT command
can be used to stop streaming.
MMON?
Example:
-00.005900
SIM960 Analog PID Controller