Appendix B:
Establishing Advanced Therapy Settings Using the Clinical Programmer
53
| SPRINT
®
PNS System – Clinician Instruction for Use [L0090-MAN-000 Rev F]
Intensity testing option using the directly set minimum and maximum intensities
Press the
button to return to directly controlling amplitude and pulse duration to change the
minimum and maximum. Tap the Save and Finish button to save your changes in the direct parameter
control. If you return to standard parameter control without saving, your changes will be discarded.
To return to using standard parameter control (intensity adjustment only), press the
icon and select
Standard Parameter Control.
Fine Resolution / Standard Resolution
Selecting
Fine Resolution will increase the resolution of the intensity adjustment step sizes (e.g.,
from step sizes of 1 to 0.5).
To return to standard resolution, press the
icon and select
Standard Resolution.
Manual Maximum / Auto Maximum
Selecting
Manual Maximum will enable you to directly set the maximum value in a similar manner
to setting the minimum value instead of utilizing the auto maximum (which automatically adds in a
buffer above the highest level of stimulation tested to provide patients with additional options during
the therapy period).
To return to auto maximum, press the
icon and select
Auto Maximum.