Chapter 4: MicroLead Placement & System Set-Up
41
| SPRINT
®
PNS System – Clinician Instruction for Use [L0090-MAN-000 Rev F]
Single-Lead
Dual-Lead
Instructions
The screen will briefly display Maximum is Set.
Dual Lead System: Continue setting stimulation
parameters for the second MicroLead (used with the
dark gray connector cables).
Setting parameters for the second (dark) MicroLead
repeats the steps used to set the parameters for the
first MicroLead.
When all settings are as desired, press the
button
to enable the Hand-Held Remote for patient use. This
will complete lead placement, and any further setting
changes will require the use of the Clinical
Programmer Tablet.
To make further adjustments to stimulation settings
press the
button.
The screen will briefly display Settings Saved &
Enabled for Patient Use. The Hand-Held Remote
screen will then change to the patient user interface.