Chapter 4: MicroLead Placement & System Set-Up
24
| SPRINT
®
PNS System – Clinician Instruction for Use [L0090-MAN-000 Rev F]
Remote Screens
Instructions
NOTE: The intent of the Fast Cycle approach is to allow for identification of
changes in muscle tension as stimulation intensity is increased. This cycling
approach is not available for procedures in which the 100 Hz frequency is
used. It is also not available for patient home use (which utilizes Slow Cycling
for 12 Hz stimulation).
If "Yes" is selected, the remote will return to the previous screen so
stimulation can be tested at the new stimulation cycle. The option to switch
to 1 Sec cycling will then be presented. Switching between Fast Cycling and
Slow Cycling can be done as many times as desired.
Press "No" to continue to the next screen.
If all settings are correct, press the
button to continue.
To erase settings and restart the process, press the
R button.
To return to testing stimulation with the same needle or lead, press the
button.
Single Lead
Dual Lead
Continue confirming placement and placing the MicroLead by following steps
in Section 4.3) Confirming Location & Placing the MicroLead(s).
Once the MicroLead has been placed and the MicroLead Connector has been
attached, press the
button to continue.
To return and adjust placement before deploying the MicroLead, press the
button.
With the lead connector attached to the lead, confirm that the lead is in the
desired location by testing stimulation again.