IM-P309-02
MI Issue 2
55
15.5.7 Density
If a mass unit in kg or t is used as flow unit (2.5.1 Volume flow QV unit), the density of the
medium must be entered in the unit of g/l. Using the entered density value, the mass flow is
calculated from the volume flow measurement.
After choosing the Density function and pressing enter, the following selection field will be
display
Density
998.2 g/l
The current density value will be displayed. According to the description in Section 14.1.3.2
"Input window / modify a value", the current value can be changed.
The value of the density is not measured. It is a parameter.
15.5.8 Volume flow LSL (information field)
This value represents the minimum lower range value based on the inside diameter of the
sensor. This value is normally set for a flow velocity of 0.25 m/s.
QV LSL
XX.XXX l/h
15.5.9 Volume flow USL (information field)
This value represents the maximum upper range value based on the inside diameter of the
sensor. This value is normally set for a flow velocity of 11 m/s.
QV USL
XX.XXX l/h