
IM-P309-02
MI Issue 2
51
15.4.4 Zero point calibration
Using the Zero point calibration function the operator can recalibrate the zero point of your
meter in the measuring system. Zero point calibration is to be realized after any installation
procedure or after any type of work has been performed on in the pipes near the sensor.
Refer also Section 11.4 Zero point calibration.
CAUTION:
This function may only be carried out if it is certain that the fluid in the
sensor is not flowing. Otherwise, the flow rates measured subsequently
will be incorrect. The sensor must be completely filled with fluid. A
partially filled sensor or air bubbles will lead to an incorrect zero point
calibration.
After choosing the Zero point calibration function and pressing enter, the current remaining
flow will be display
0.00 l/h
cal.? (no)
According to the description in Section 14.1.3.2 Selection window / make a selection, "yes" or
"no" can be selected. By pressing Esc or toggling to [no] the operator can cancel the current
action. Enter [yes] to have the zero point recalibrated.
15.4.5 Filter
For a noise reduction of the actual readings a signal filter can be used. Following settings
are available:
• none
• weak
• mid (standard factory setting)
• strong
Selecting "weak" or "mid" influences the dynamics of the actual readings not or just very
less. The setting of damping time (see section 8.11.6 "Damping") determines the dynamic
behaviour of sensor and transmitter. Filter "strong" operates as an intense low pass filter
to reduce the noise of the actual readings. Then if setting the damping time shorter than 3
seconds, the low pass filter determines the dynamic behavior to actual reading’s variations.
According to the description in Section 14.1.3.2 Selection window / make a selection, noise
filter type can be selected.
Summary of Contents for ELM
Page 2: ...IM P309 02 MI Issue 2 2 ...
Page 78: ...IM P309 02 MI Issue 2 78 ...
Page 79: ...IM P309 02 MI Issue 2 79 ...
Page 80: ...IM P309 02 MI Issue 2 80 ...