SS1000 Operator’s Manual
3
–
0
1
-
9
2
-
4
A
.
v
e
r
9
7
0
2
0
-
0
0
0
9
4
4
1
3.
Press the
0
key followed by the
*
key to turn peak tracking off. Press
the
#
key followed by the
1
key to return to
Mode 1
.
4.
Press the
#
key followed by the
2
key. The LCD prompts for a
numeric password.
5.
Enter the user password (
7758
) on the keypad, then press the
*
key
to enter the number. This puts you into the factory mode.
6.
Enter the
Phase
value (as provided with the calibration documents)
for the desired range followed by the
*
key.
7.
Enter the
Midpoint
value (as provided with the calibration
documents) for the desired range followed by the
*
key.
8.
Repeat for
Ramp Amplitude
,
Imod
,
RCalb
,
Zero Offset
,
Span
,
PA1
,
PA2
,
PA3
and
PA4
.
9.
Press the
#
key followed by the
1
key to return to
Mode 1
.
10.
Follow the procedure under
“To change parameters in Mode 2”
on page 3-6 to enter new parameter values.
When changing parameters in factory mode, be careful not to
change any parameters other than those specified for range
change or the calibration and/or operation of the analyzer may be
adversely affected.
<SET PARAMETER MODE>
Enter password
v2.40 08-12-08 DM
<SET PARAMETER MODE>
Phase
320
Enter a value
<SET PARAMETER MODE>
Midpoint
59.70000
Enter a value