Logger Rate
For applications where an external data logger is employed, use the logging
rate to set the number of measurements to include in the running average. The
display and the current loop output will each have a value representing the
running average of the concentration over a number of measurements equal to
Logger Rate
.
Peak Tracking
The peak tracking function is a software utility that continuously adjusts the
laser current to keep the absorption peak of the measured component at the
center of the scan. There are three choices:
0
for no peak tracking,
1
for peak
tracking (default), or
2
to reset the peak tracking function. In most cases, the
peak tracking should be left on (i.e.
Peak Tracking
set to
1
).
Temperature Unit
The
Temperature Unit
parameter designates the display units for the
measured cell temperature. There are two choices:
0
for degrees Celsius and
1
for Fahrenheit. The default value is the standard unit of measurement in the
region the analyzer is being used.
<SET PARAMETER MODE>
Logger Rate
4
Enter a value
<SET PARAMETER MODE>
Peak Tracking
1
0:Off 1:Track 2:Reset
<SET PARAMETER MODE>
Temperature Unit
0
0:C 1:F