Measurement and Control Parameters Defined
S Factor
The
S Factor
parameter is a user definable value that enables adjustment
(without affecting the factory calibration) of the analyzer response (or slope)
in the field [see
“Adjusting Analyzer Reading to Match Specific
Standard(s)”
on page 3-12].
S Factor Offset
The
S Factor Offset
parameter is a user definable value that enables
adjustment (without affecting the factory calibration) of the analyzer offset in
the field [see
“Adjusting Analyzer Reading to Match Specific
Standard(s)”
on page 3-12].
# Spectrum Average
The
# Spectrum Average
parameter sets the number of scans that the
analyzer averages for when determining the concentration. Averaging over
multiple scans lowers noise in the measurement but inherently increases the
response time. Each scan adds about 0.25 seconds to the response time. For
example, if
#Spectrum Average
is set to “4,” an updated concentration value
will be calculated about once every second.
<SET PARAMETER MODE>
S Factor
1.00000
Enter a value
<SET PARAMETER MODE>
S Factor Offset
0.00000
Enter a value (%)
<SET PARAMETER MODE>
# Spectrum Average
16
Enter a value