Spectra Precision FOCUS 30 User Guide 57
4 Setup
The Laser Pointer
The Spectra Precision FOCUS 30 Total Station uses a red laser beam to measure and
as a Laser Pointer. The Laser Pointer is coaxial with the line of sight of the telescope.
If the instrument is well adjusted, the red Laser Pointer coincides with the line of
sight. External influences such as shock or large temperature fluctuations can displace
the red Laser Pointer relative to the line of sight.
Aligning the Laser Pointer
C
Caution –
Viewing the laser spot on the adjustment target through the telescope is safe.
C
Warning –
Do not try to make the adjustment using a prism, see
C
Warning –
Do not use the Laser Pointer as an aid when searching for prisms, see
.
To avoid faulty measurements using the Laser Pointer, use the supplied adjustment
target to check the laser alignment regularly and before you attempt precise distance
measurements:
1.
Setup the adjustment target 10-15 meters away, facing the instrument.
2.
Aim the instrument to the center of the target plate and then inspect the position
of the red laser spot in relation to the telescope cross-hairs.
3.
If the red laser spot lies outside the cross-hairs, adjust the direction of the beam
until it matches the cross-hairs, see
.
Fig. 4-46 Laser adjustment plate (71199035)
Summary of Contents for Focus 30
Page 1: ...Spectra preciSion 30 TOTAL STATION USER GUIDE...
Page 2: ......
Page 11: ...Spectra Precision FOCUS 30 User Guide IX Declaration of Conformity...
Page 12: ...Declaration of Conformity X Spectra Precision FOCUS 30 User Guide...
Page 16: ...Table of Contents Spectra Precision FOCUS 30 User Guide P N 77701035 Version 1 02...
Page 26: ...2 Inspection Care and Maintenance 10 Spectra Precision FOCUS 30 User Guide...
Page 84: ...4 Setup 68 Spectra Precision FOCUS 30 User Guide...
Page 88: ...5 Instrument Operation Methods 72 Spectra Precision FOCUS 30 User Guide...
Page 125: ......