Setup
91369 Service Manual
2-32
• QRS/SPO2 TONE ENABLE / ALWAYS/DURING ALARM — Controls whether the monitor sounds the
QRS or SpO
2
tone all the time (ALWAYS) or only during alarm conditions (DURING ALARM). The QRS or
SpO
2
tone must also be enabled using the controls within those parameters’ menus.
Alarm Watch Setup
The ALARM WATCH SETUP key controls how the monitor responds to alarm watch messages received from
other monitors.
• ROTATE ALARM WATCH / ON/OFF — This key enables (ON) and disables (OFF) the alarm watch
rotation feature (default is OFF).
• ALARM WATCH ROTATION / PRIORITY/SIMPLE — When a monitor receives more than one alarm
watch message at a time, it uses a “first-in, first-out” rotation scheme to display the alarming parameters.
This key is only enabled when ROTATE ALARM WATCH is enabled. The following choices of rotation
schemes are available:
-
PRIORITY rotation cycles through the alarming parameters based on each parameter’s alarm priority
(for example, parameters with high-priority alarms display before parameters with medium- or low-
priority alarms).
-
SIMPLE rotation cycles through the alarming parameters in the order that they go into alarm (first
come, first served).
• ROTATION TIME / 15 SEC/30 SEC — You can choose to display an alarm watch for either 15 or
30 seconds before cycling to the next alarm watch if there are two or more alarm-watched beds in alarm.
This key is only enabled when ROTATE ALARM WATCH is enabled.
Change Clinical Password
The CHANGE CLINICAL PASSWORD key enables you to change the password used to access the
Clinical
Level
menu.
To change the clinical password:
1 Enter the current clinical password in the
Password
field using the on-screen keyboard (passwords are not
case-sensitive).
2 Enter the new clinical password in the
New Password
field and enter the same password again in the
Verify Password
field using the on-screen keyboard.
Note:
If the clinical password is forgotten, contact your system administrator.
Reset Monitor
The RESET MONITOR key allows you to reboot the monitor after changing settings for the following items (the
monitor must be rebooted before these changes can take effect):
• Subnet access
• Units of measurement
Touching the RESET MONITOR key displays the
Reset Monitor
dialog box. Select
Reset Monitor
to proceed
or
Cancel Reset
to cancel.