91369 Service Manual
2-29
Setup
Vitalink
Vitalink is a Draeger-defined asynchronous serial communication protocol for use only with Draeger devices.
To set up
Vitalink
:
1 Touch MONITOR SETUP.
2 Touch PRIVILEGED ACCESS.
3 Enter the biomed password (default is
biomed
).
4 Touch SERIAL PORTS.
5 Touch ASSIGNMENT.
6 Touch VITALINK.
7 Touch PREVIOUS MENU.
Serial Settings
To set serial settings:
1 Touch SETTINGS.
2 Touch the desired setting key(s) to display and set the desired settings.
3 Touch NORMAL SCREEN to effect changes.
Monitor Calibration
The MONITOR CALIBRATION key in the
Biomed Level
menu allows you to perform a touchscreen calibration
in the event the touchscreen becomes difficult to use or a replacement has been installed. Refer to
on page 4-4 for instructions on performing this calibration.
Change Biomed Password
The CHANGE BIOMED PASSWORD key in the
Biomed Level
menu enables you to change the password
used to access the
Biomed Level
menu.
To change the biomed password:
1 Enter the current biomed password in the
Password
field using the on-screen keyboard (passwords are not
case-sensitive).
2 Enter the new biomed password in the
New Password
field and enter the same password again in the
Verify Password
field using the on-screen keyboard.
Note:
If the biomed password is forgotten, contact your Spacelabs Healthcare Field Service Engineer.