91369 Service Manual
2-13
Setup
Module Tests
To verify that the monitor functions correctly with parameter modules:
1 Insert an ECG module without the patient cables connected. Verify that the ECG parameter key is
displayed.
2 Connect a patient simulator to the ECG input with a 5-lead patient cable, and set the simulator to a known
rate.
•
Verify that the heart rate and lead being monitored are displayed to the right of the ECG parameter
key.
•
Verify that the ECG waveform is displayed.
3 Disconnect the patient cable. After 10 seconds, verify that the
LEADS OFF
message appears, the
parameter key flashes, and an alarm tone sounds.
4 Reconnect the patient cable and verify that the
LEADS OFF
message clears and the alarm stops.
5 Connect a patient simulator to the invasive pressure inputs.
6 Zero the pressures and verify that the numerics and waveforms are accurate.
7 Verify that the key tone sounds each time a key is selected.
Configuring the Monitor
The
Biomed Level
menu displays when the biomed password (default is
biomed
) is entered into the
Privileged Access
window. Refer to
on page 7-1 for the
Biomed Level
menu structure.
Network Setup
Note:
The NETWORK SETUP key only displays on monitors that are configured for network operation.
Touch NETWORK SETUP to display the
Monitor Setup - Network Configuration
window. This window
contains an on-screen keyboard and three tabs:
TCP/IP
,
Monitor
,
and
Printers
. Proper network operation
requires that each device on the network have a unique network address, monitor ID, and monitor name. If the
wireless option has been installed on the monitor, three more tabs will be present:
WLAN
,
Security
, and
Advanced
.
Editing Tab Fields
The fields within a tab on the
Monitor Setup - Network Configuration
window can be edited by selecting the
field and entering new information using the on-screen or optional external keyboard.
When editing, adding, or deleting, press ENTER or TAB to cycle to the next input cell. Any changed or added
items are stored in the monitor’s non-volatile memory when SAVE is selected. The description of each tab
indicates when that change takes effect (for example, immediately or after a monitor reset occurs).
To edit text within a tab:
1 Select an item from the list.
-OR-