11
Operation cont.
1. Set the rotary function switch to
the position
2. Press the
MODE
button until
the “
nF
” symbol appears on
the display.
3. Touch the test lead probes to the
capacitor under test.
4. Read the capacitance on the LCD
display. It may take up to a minute
to get a stable reading on a large capacitor.
Capacitance Measurements
Safely discharge capacitors before
taking capacitance measurements.
WARNING:
1. Set the rotary function switch to
the
position.
2. Press the
MODE
button until
the “ ” symbol appears on the
LCD display.
3. Touch the test lead probes to the
device or wire under test.
4. A beeper will sound if the resistance
is approx.100 ohms or less and
the resistance reading will be
shown on the LCD display.
Continuity
WIRE
Never test continuity on a live circuit.
WARNING:
Digital Multimeter
16020N
REL
Digital Multimeter
16020N
REL
Digital Multimeter
16020N
REL
Digital Multimeter
16020N
REL
12
Operation cont.
1. The non-contact voltage detector
operates when the meter is set to
any measuring function. The detector
does not operate when Auto Power
Off turns the meter off or when the
rotary function switch is set to the
off position.
2. Hold the detector close to the AC
voltage being tested.
3. If AC voltage within the specified
range is present, the indicator
light will illuminate.
Non-contact AC Voltage Detection: 100V to 600V
Risk of Electrocution. Before use, always test the
Voltage Detector on a known live circuit to verify proper operation.
WARNING:
NOTE:
Insulation type and thickness, and distance from the voltage source
and other factors may effect operation. Use other methods to verify live voltage
if there is any uncertainty.
NOTE:
The detector is designed with high sensitivity. Static electricity or other
sources of energy may randomly activate the sensor. This is normal operation.
NOTE:
The detector only activates the indicator light when AC voltage is
present. It does not indicate the voltage level on the LCD display.
Digital
Multimet
er
16020N
REL