ST2523 Series Operation Manual Ver1.2
51
Query syntax: BINSETup: BINA N?
Return value: <NR3><NL^END>
<NR3> the upper and lower limit of the primary parameter
The:BINSETup:BINB command set the upper and lower limit of the secondary parameter. According to
different modes, the upper and lower limit of absolute value or percentage can be set. The:BINSETup:BINB?
query returns the current limit of current mode.
Command syntax: BINSETup:BINB N:<value,value>
N Bin number, ranges from 1~9.
<value,value> the upper and lower limit of the secondary parameter
For example: WrtCmd(“BINSETup: BINB 2:15,10”);set the upper limit value of secondary parameter in
BIN2 as 15 and the lower limit as 10.
Query syntax: BINSETup: BINB N?
Return value: <NR3><NL^END>
<NR3> the upper and lower limit of the secondary parameter